Chemistry of Solutions 202-SN2-RE • CEGEP, Montreal

Corrected exercises: measurement, uncertainty and graphical analysis of data (202-SN2-RE)

This is the corrected exercise set on measurement, uncertainty and the graphical treatment of data in the laboratory for Chemistry of Solutions, 202-SN2-RE (formerly 202-NYB). It is the experimental side of the course outline, the one that is assessed in every lab report without ever getting a whole lecture to itself. Part A builds the toolkit in the order the lab asks for it: volumetric glassware and its tolerance, how uncertainties accumulate in a serial dilution, the regression line calculated by hand on a calibration curve, the coefficient of determination, the uncertainty on a slope by the extreme lines, and reading an unknown concentration backwards off the curve. Part B moves up to test level: the plot that tells first order from second order, propagation through a logarithm, the exponent that multiplies a relative uncertainty, an iron tablet assayed by spectrophotometry from the flask to the waste container, and an activation energy taken from an Arrhenius slope.

It is written for CEGEP students in Science (Sciences de la nature), under the revised 202-SN2-RE course outline as under the former 202-NYB. It assumes the measurement chapter of general chemistry, significant figures and the uncertainty of a single measurement, and picks up exactly where that chapter stops.

The thread of the set fits in one sentence: in the chemistry of solutions, the constant you are after is almost never a reading, it is a SLOPE. Everything else follows, including the least intuitive conclusion of the chapter, namely that a slope is blind to anything that shifts every point by the same amount, and that its uncertainty can be read neither off a single point nor off the coefficient of determination.

10 corrected exercises • 100 points • 150 minutes

This chapter is part of 202-SN2-RE, every chapter with its set and sheet Before you start Revision sheet: the traps and the method for this chapter

Self-checking set Type your answers and check them question by question: the page tells you right or wrong before you open the solution. With an account, every answer right on the first try earns points, and your right answers stay filled in when you come back.

Course recap

  • • Relative uncertainty =absolute uncertaintymeasured value=\dfrac{\text{absolute uncertainty}}{\text{measured value}}. Product or quotient: the RELATIVE uncertainties add. Sum or difference: the ABSOLUTE ones add. Power nn: the relative uncertainty is MULTIPLIED by nn, and a square root halves it.
  • • Class A glassware, typical tolerances: 100.00100.00 mL volumetric flask ±0.08\pm 0.08 mL, 250.0250.0 mL flask ±0.15\pm 0.15 mL, 10.0010.00 mL volumetric pipette ±0.02\pm 0.02 mL, 5.005.00 mL pipette ±0.015\pm 0.015 mL, burette ±0.05\pm 0.05 mL PER READING, 100100 mL graduated cylinder ±1\pm 1 mL.
  • • Serial dilution: the dilution factors MULTIPLY, and the relative uncertainties ADD. One more dilution is one more uncertainty, even when it is carried out perfectly.
  • • Beer-Lambert law: A=εℓCA=\varepsilon \ell C, a straight line through the origin when the blank is done properly. It stops being linear above A≈1A\approx 1, and a calibration curve is valid only over the interval where it was established.
  • • Linear regression: a=SxySxxa=\dfrac{S_{xy}}{S_{xx}} with Sxx=∑(xi−xˉ)2S_{xx}=\sum (x_{i}-\bar x)^{2} and Sxy=∑(xi−xˉ)(yi−yˉ)S_{xy}=\sum (x_{i}-\bar x)(y_{i}-\bar y), then b=yˉ−axˉb=\bar y-a\bar x. The line ALWAYS passes through the mean point (xˉ,yˉ)(\bar x, \bar y).
  • • Coefficient of determination: r2=Sxy2SxxSyyr^{2}=\dfrac{S_{xy}^{2}}{S_{xx}S_{yy}}. It measures how well the points LINE UP, not whether the model is right, and it supplies no uncertainty.
  • • Extreme lines: the steepest and the least steep line that still pass through every error bar. They give a=amax+amin2a=\dfrac{a_{max}+a_{min}}{2} and Δa=amax−amin2\Delta a=\dfrac{a_{max}-a_{min}}{2}, with no statistics at all.
  • • The four linearizations of the course: first order, ln⁡[A]\ln[A] against tt, slope −k-k; second order, 1[A]\dfrac{1}{[A]} against tt, slope +k+k; Arrhenius, ln⁡k\ln k against 1T\dfrac{1}{T}, slope −EaR-\dfrac{E_{a}}{R}; Beer-Lambert, AA against CC, slope εℓ\varepsilon\ell.
  • • Passing through a logarithm: Δ[H3O+][H3O+]=ln⁡(10)×ΔpH≈2.30 ΔpH\dfrac{\Delta[H_{3}O^{+}]}{[H_{3}O^{+}]}=\ln(10)\times\Delta\mathrm{pH}\approx 2.30\,\Delta\mathrm{pH}. A ΔpH\Delta\mathrm{pH} of 0.020.02, which looks negligible, is worth 4.6 %4.6\,\% on the concentration.
  • • Significant figures of a logarithmic result: only the digits after the decimal point (the mantissa) count. pH=2.87\mathrm{pH}=2.87 carries two significant figures, so [H3O+]=1.3×10−3[H_{3}O^{+}]=1.3\times 10^{-3} mol/L, not 1.3490×10−31.3490\times 10^{-3}.
  • • Relative deviation from a reference value: ∣xmeas−xref∣xref×100\dfrac{|x_{meas}-x_{ref}|}{x_{ref}}\times 100. The measurement is compatible if the deviation fits within the uncertainty, and it is quantified as a NUMBER of uncertainties.
  • • Random error: it scatters, it shrinks with repetition, it widens the extreme lines. Systematic error: it shifts, it NEVER shrinks with repetition, and only a calibration or a different procedure corrects it.
  • • An uncertainty is rounded to one significant figure, and the result lines up on it: C=(1.6±0.1)C=(1.6\pm 0.1) mg/L.
  • • Health, safety and environment: safety glasses and lab coat at all times, safety data sheet read before the experiment, heavy-metal waste, solvents and acidic solutions into the designated containers, never down the sink.

Part A: from glassware to the regression line (/50)

Exercise 1: Volumetric glassware and serial dilution

The five standards of a calibration curve are to be prepared from an iron(II) stock solution of C0=1.000×10−2C_{0}=1.000\times 10^{-2} mol/L, known to within 0.10 %0.10\,\%. Class A tolerances stated by the manufacturer: 100.00100.00 mL volumetric flask ±0.08\pm 0.08 mL, 10.0010.00 mL volumetric pipette ±0.02\pm 0.02 mL, 100100 mL graduated cylinder ±1\pm 1 mL. The figure adds the burette, whose TWO readings at ±0.05\pm 0.05 mL give the uncertainty of ±0.10\pm 0.10 mL on the 25.0025.00 mL delivered.

A serial dilution consists of taking 10.0010.00 mL with the pipette and making up to 100.00100.00 mL in the flask, then starting again on the solution obtained.

volumetric flask100.00 mL± 0.08 mLvolumetric pipette10.00 mL± 0.02 mLcylinder100 mL± 1 mLburette25.00 mL delivered± 0.10 mL2 readings at ± 0.05 mL
  • a) Calculate the relative uncertainty, as a percentage, of the 10.0010.00 mL volumetric pipette, of the 100.00100.00 mL volumetric flask and of the 100100 mL cylinder.
  • b) Calculate the relative uncertainty on the concentration obtained after ONE dilution, taking the stock solution to be perfect for this question.
  • c) Three identical dilutions are performed in a row. Give the final concentration, the total relative uncertainty (stock solution included) and the absolute uncertainty.
  • d) A classmate replaces the volumetric pipette with a 1010 mL cylinder at ±0.1\pm 0.1 mL, for all three steps. What happens to the total relative uncertainty, and by what factor?
  • e) The five standards prepared this way supply the X-AXIS of the calibration curve. Explain why a miscalibrated pipette is, here, far more serious than an imprecise one.

Type your answers, the page tells you right or wrong 0/11

a)
b)
c)
d)
e)
Show the solution

Answers

  • a) 0.200 %0.200\,\% for the pipette, 0.080 %0.080\,\% for the flask, 1.0 %1.0\,\% for the cylinder
  • b) 0.280 %0.280\,\%
  • c) C3=(1.000±0.009)×10−5C_{3}=(1.000\pm 0.009)\times 10^{-5} mol/L, that is 0.94 %0.94\,\%
  • d) 3.34 %3.34\,\%, that is 3.63.6 times more
  • e) It shifts every standard the same way: the slope tilts and nothing signals it
a)

Pipette: 0.0210.00×100=0.200 %\dfrac{0.02}{10.00}\times 100=0.200\,\%. Flask: 0.08100.00×100=0.080 %\dfrac{0.08}{100.00}\times 100=0.080\,\%. Cylinder: 1100×100=1.0 %\dfrac{1}{100}\times 100=1.0\,\%. The cylinder is five times worse than the pipette although it measures ten times more: a tolerance can only be compared once it is expressed relative to the volume measured.

b)

The diluted concentration is C1=C0VpipetteVflaskC_{1}=C_{0}\dfrac{V_{pipette}}{V_{flask}}: a quotient, so the RELATIVE uncertainties add. 0.200+0.080=0.280 %0.200+0.080=0.280\,\%. Note that the two instruments do not weigh the same: the pipette alone supplies five sevenths of the total, because it measures the smallest volume in the chain.

c)

Each dilution divides by 1010, so C3=C01000=1.000×10−5C_{3}=\dfrac{C_{0}}{1000}=1.000\times 10^{-5} mol/L. The three steps are three successive quotients: 3×0.280=0.840 %3\times 0.280=0.840\,\%, plus the 0.10 %0.10\,\% of the stock solution, giving 0.94 %0.94\,\%. Absolute uncertainty: 0.0094×1.000×10−5=9.4×10−80.0094\times 1.000\times 10^{-5}=9.4\times 10^{-8} mol/L, so C3=(1.000±0.009)×10−5C_{3}=(1.000\pm 0.009)\times 10^{-5} mol/L. The lesson of the calculation: dilution factors MULTIPLY but uncertainties ADD, so one more dilution always costs something, even when it is carried out perfectly.

d)

One step rises to 0.110.0×100+0.080=1.08 %\dfrac{0.1}{10.0}\times 100+0.080=1.08\,\%, the three steps to 3.24 %3.24\,\%, and the total to 3.34 %3.34\,\% with the stock solution. The factor is 3.340.94=3.6\dfrac{3.34}{0.94}=3.6: the uncertainty on every standard is multiplied by more than three, and with it the uncertainty on the slope that will be drawn from them. This is no saving of time, it is a net loss of precision on the final result, about half a significant figure.

e)

Because the two defects do not do the same damage to a line. An IMPRECISE pipette delivers 9.989.98 mL one time and 10.0210.02 mL the next: the errors change sign from one standard to the next, the points scatter around the line, and that scatter is visible, it widens the extreme lines and so it shows up in the published uncertainty. A miscalibrated pipette always delivers 9.909.90 mL: all five standards are 1 %1\,\% too dilute, in the same direction, the points stay perfectly aligned, the coefficient of determination stays superb, and the whole line tilts by 1 %1\,\%. The systematic error therefore hides exactly where nobody looks, in the SLOPE, while every quality indicator stays green. Only an independent check, weighing the water delivered by the pipette, reveals it.

Tick the exercises you have done or want to review: a free account, no password, keeps your ticks and your right answers from one visit to the next, tells you which chapter to tackle next and lets you ask for the exercise you are missing. Create your space, an email is enough.

Exercise 2: Linear regression by hand, from the scatter plot to the line

The five standards of iron(II) complexed with o-phenanthroline are run in the spectrophotometer at 510510 nm, in a 1.001.00 cm cuvette. The readings are: C=0.500C=0.500 mg/L, A=0.096A=0.096; C=1.000C=1.000 mg/L, A=0.190A=0.190; C=1.500C=1.500 mg/L, A=0.290A=0.290; C=2.000C=2.000 mg/L, A=0.382A=0.382; C=2.500C=2.500 mg/L, A=0.478A=0.478.

We look for the line A=aC+bA=aC+b by the method of least squares, with a=SCASCCa=\dfrac{S_{CA}}{S_{CC}}, SCC=∑(Ci−Cˉ)2S_{CC}=\sum (C_{i}-\bar C)^{2}, SCA=∑(Ci−Cˉ)(Ai−Aˉ)S_{CA}=\sum (C_{i}-\bar C)(A_{i}-\bar A) and b=Aˉ−aCˉb=\bar A-a\bar C.

0.511.522.530.10.20.30.40.50.6C (mg/L)A
  • a) Calculate Cˉ\bar C and Aˉ\bar A, and say what is special about this point for the line we are looking for.
  • b) Calculate SCCS_{CC} and SCAS_{CA}, then the slope. Give its unit and its physical meaning.
  • c) Calculate the y-intercept. What should it be, and what would a value of 0.0400.040 have signalled?
  • d) Calculate the coefficient of determination r2=SCA2SCCSAAr^{2}=\dfrac{S_{CA}^{2}}{S_{CC}S_{AA}}, given that SAA=0.091405S_{AA}=0.091405.
  • e) A classmate concludes: “r2=0.9999r^{2}=0.9999, so the Beer-Lambert law is proven and my slope is exact to the fourth decimal place.” There are two mistakes. Which?

Type your answers, the page tells you right or wrong 0/10

a)
b)
c)
d)
e)
Show the solution

Answers

  • a) Cˉ=1.500\bar C=1.500 mg/L and Aˉ=0.2872\bar A=0.2872: the line always passes through this point
  • b) SCC=2.50S_{CC}=2.50, SCA=0.4780S_{CA}=0.4780, slope a=0.1912a=0.1912 L/(mg⋅\cdotcm)
  • c) b=0.0004b=0.0004, so zero within measurement error; 0.0400.040 would have signalled a failed blank
  • d) r2=0.9999r^{2}=0.9999
  • e) It measures alignment, not the model, and it gives no uncertainty on the slope
a)

Cˉ=0.500+1.000+1.500+2.000+2.5005=1.500\bar C=\dfrac{0.500+1.000+1.500+2.000+2.500}{5}=1.500 mg/L and Aˉ=1.4365=0.2872\bar A=\dfrac{1.436}{5}=0.2872. This point has a property that costs nothing and serves as a permanent check: the least-squares line ALWAYS passes through the mean point, however scattered the data. If your line drawn with a ruler does not pass through (1.500,0.2872)(1.500, 0.2872), it is wrong before any calculation.

b)

The deviations in CC are −1.00-1.00, −0.50-0.50, 00, +0.50+0.50, +1.00+1.00, so SCC=1+0.25+0+0.25+1=2.50S_{CC}=1+0.25+0+0.25+1=2.50. For SCAS_{CA}: (−1.00)(−0.1912)+(−0.50)(−0.0972)+0+(0.50)(0.0948)+(1.00)(0.1908)=0.1912+0.0486+0+0.0474+0.1908=0.4780(-1.00)(-0.1912)+(-0.50)(-0.0972)+0+(0.50)(0.0948)+(1.00)(0.1908)=0.1912+0.0486+0+0.0474+0.1908=0.4780. Hence a=0.47802.50=0.1912a=\dfrac{0.4780}{2.50}=0.1912 L/(mg⋅\cdotcm) for a 1.001.00 cm cuvette. Physically, this slope is the product εℓ\varepsilon \ell of Beer-Lambert: it, and it alone, holds the chemical information of the calibration.

c)

b=Aˉ−aCˉ=0.2872−0.1912×1.500=0.2872−0.2868=0.0004b=\bar A-a\bar C=0.2872-0.1912\times 1.500=0.2872-0.2868=0.0004. It should be zero: at zero concentration the cuvette holds the blank, and the spectrophotometer was set to zero on that very blank. A y-intercept of 0.0400.040 would have meant that every absorbance is too high by 0.0400.040, so a failed blank, a scratched cuvette or a reagent that is coloured by itself. It is the cheapest check of the session, and it detects only one thing, but precisely the one that escapes the coefficient of determination: a SHIFT common to all the points.

d)

r2=(0.4780)22.50×0.091405=0.2284840.228513=0.9999r^{2}=\dfrac{(0.4780)^{2}}{2.50\times 0.091405}=\dfrac{0.228484}{0.228513}=0.9999. The points are therefore practically aligned.

e)

First mistake: r2r^{2} measures how well the points LINE UP, not whether the model is right. A slightly curved line over a narrow interval also gives 0.9990.999, and a perfect systematic error gives exactly 1.0001.000 while being entirely wrong. An r2r^{2} is never read alone: it is compared with that of the competing model, or backed up by an examination of the residuals. Second mistake: r2r^{2} contains NO uncertainty. It says the points are aligned, it does not say by how much the slope could move if the experiment were repeated. The four digits the calculator displays for the slope are the result of a division, not a measurement: a separate method is needed, the extreme lines of the next exercise, to know how many of those digits survive.

0.511.522.530.10.20.30.40.50.6A = 0.1912 C + 0.0004r² = 0.9999C (mg/L)A

Exercise 3: The uncertainty on the slope by the extreme lines

The laboratory spectrophotometer is rated at ±0.008\pm 0.008 absorbance units, with the cuvette removed and replaced between readings. Each point of the previous calibration therefore carries a vertical bar of ±0.008\pm 0.008, drawn on the figure; the uncertainty on the concentrations of the standards is a hundred times smaller and is neglected here.

The extreme-lines method consists of drawing with a ruler the STEEPEST line and the LEAST steep line that still pass through every bar, then setting a=amax+amin2a=\dfrac{a_{max}+a_{min}}{2} and Δa=amax−amin2\Delta a=\dfrac{a_{max}-a_{min}}{2}. No statistics are involved.

0.511.522.530.10.20.30.40.50.6C (mg/L)A
  • a) Through which points must the steepest line pass, and why those two rather than the other three?
  • b) Calculate the maximum slope.
  • c) Calculate the minimum slope.
  • d) Give the slope with its uncertainty, then the relative uncertainty. How many digits of the value 0.19120.1912 found by regression survive?
  • e) Calculate the two extreme y-intercepts. What can be asserted about the blank, and what would have to be concluded if the interval obtained had been [0.030,0.055][0.030, 0.055]?

Type your answers, the page tells you right or wrong 0/10

a)
b)
c)
d)
e)
Show the solution

Answers

  • a) Through (0.500,0.088)(0.500, 0.088) and (2.500,0.486)(2.500, 0.486), the two end points
  • b) amax=0.199a_{max}=0.199 L/(mg⋅\cdotcm)
  • c) amin=0.183a_{min}=0.183 L/(mg⋅\cdotcm)
  • d) a=(0.191±0.008)a=(0.191\pm 0.008) L/(mg⋅\cdotcm), that is 4.2 %4.2\,\%: two digits survive
  • e) b∈[−0.012,+0.013]b\in[-0.012, +0.013], compatible with zero; [0.030,0.055][0.030, 0.055] would signal a systematic error
a)

Through the BOTTOM of the bar of the first point and the TOP of the last: (0.500,0.088)(0.500, 0.088) and (2.500,0.486)(2.500, 0.486). Two reasons. First, a slope is a ratio of differences: to make it as large as possible, you lower the left-hand point and raise the right-hand one, and it is the end points that give the longest lever arm. Second, the three middle points constrain nothing here, since their bars are crossed by the line obtained; they would have to be checked, and the line moved if one of them were missed, which happens as soon as a point is an outlier.

b)

amax=0.486−0.0882.500−0.500=0.3982.000=0.199a_{max}=\dfrac{0.486-0.088}{2.500-0.500}=\dfrac{0.398}{2.000}=0.199 L/(mg⋅\cdotcm).

c)

Symmetrically, through the top of the first point and the bottom of the last: amin=0.470−0.1042.000=0.3662.000=0.183a_{min}=\dfrac{0.470-0.104}{2.000}=\dfrac{0.366}{2.000}=0.183 L/(mg⋅\cdotcm).

d)

a=0.199+0.1832=0.191a=\dfrac{0.199+0.183}{2}=0.191 and Δa=0.199−0.1832=0.008\Delta a=\dfrac{0.199-0.183}{2}=0.008, that is a=(0.191±0.008)a=(0.191\pm 0.008) L/(mg⋅\cdotcm) and 0.0080.191×100=4.2 %\dfrac{0.008}{0.191}\times 100=4.2\,\%. The midpoint of the two extreme lines lands on 0.1910.191, within a thousandth of the regression: the two methods agree on the VALUE. They do not agree on the number of digits. Of the four digits of 0.19120.1912, TWO survive: the uncertainty already falls on the third. Writing 0.19120.1912 in a report claims a precision about eighty times better than that of the instrument.

e)

Upper line: bmax=0.088−0.199×0.500=−0.0115b_{max}=0.088-0.199\times 0.500=-0.0115. Lower line: bmin=0.104−0.183×0.500=+0.0125b_{min}=0.104-0.183\times 0.500=+0.0125. The interval [−0.012,+0.013][-0.012, +0.013] contains zero: the blank is compatible with zero, nothing contradicts it, and that is all you can assert. An interval [0.030,0.055][0.030, 0.055], on the other hand, would exclude zero: there would then be a residual absorbance of a few hundredths that the scatter of the measurements cannot explain, hence a SYSTEMATIC error, a scratched cuvette, a blank prepared without the reagent or a slightly coloured reagent. And that error, it must be stressed, would leave the slope untouched: it raises every point by the same amount, which moves bb without touching aa. This is the deep reason why a constant is sought in a slope rather than in an isolated reading.

0.511.522.530.10.20.30.40.50.6upper line: slope 0.199lower line: slope 0.183C (mg/L)A

Exercise 4: Reading the calibration curve backwards, and its uncertainty

We now have the calibration line A=aC+bA=aC+b with a=(0.191±0.008)a=(0.191\pm 0.008) L/(mg⋅\cdotcm) and b=0.000b=0.000, established for CC between 0.5000.500 and 2.5002.500 mg/L. A river water sample is treated with the same reagent: 5.005.00 mL is taken with the pipette (±0.015\pm 0.015 mL) and made up to 100.00100.00 mL (±0.08\pm 0.08 mL), and the cuvette gives A=0.305±0.008A=0.305\pm 0.008.

The diagram below shows how the sample is prepared.

riverwaterpipette5.00 mL± 0.015 mLflask100.00 mL± 0.08 mLcuvetteA = 0.305± 0.008
  • a) Calculate the iron concentration of the solution that went into the spectrophotometer.
  • b) Calculate its relative and then absolute uncertainty, and write the publishable result.
  • c) Work back to the iron concentration of the river water, with its uncertainty.
  • d) A second river gives A=1.20A=1.20. A classmate applies the same line and announces 6.276.27 mg/L. Why is this number meaningless, and what should be done?
  • e) A third gives A=0.012A=0.012. Calculate the relative uncertainty on the absorbance and conclude on how a sample must be brought in front of the instrument.

Type your answers, the page tells you right or wrong 0/10

a)
b)
c)
d)
e)
Show the solution

Answers

  • a) C=1.596C=1.596 mg/L before looking at the uncertainty
  • b) 6.8 %6.8\,\%, that is C=(1.6±0.1)C=(1.6\pm 0.1) mg/L; the slope dominates
  • c) C=(32±2)C=(32\pm 2) mg/L, that is 7.2 %7.2\,\%
  • d) Above A≈1A\approx 1 the law is no longer linear: dilute and read again
  • e) 67 %67\,\%: bring the sample to the middle of the range, around A=0.3A=0.3 to 0.70.7
a)

Invert the line: C=A−ba=0.305−0.0000.191=1.596C=\dfrac{A-b}{a}=\dfrac{0.305-0.000}{0.191}=1.596 mg/L, which the calculator will give to as many digits as you like. It is the easiest calculation of the chapter, and also the one that is spoiled by stopping there.

b)

The quotient Aa\dfrac{A}{a} makes the RELATIVE uncertainties add: 0.0080.305×100=2.6 %\dfrac{0.008}{0.305}\times 100=2.6\,\% for the absorbance, 4.2 %4.2\,\% for the slope, total 6.8 %6.8\,\%. Absolute uncertainty: 0.068×1.596=0.110.068\times 1.596=0.11 mg/L, rounded to one significant figure, so 0.10.1 mg/L. Publishable result: C=(1.6±0.1)C=(1.6\pm 0.1) mg/L. Two digits, not four. Note which of the two contributions dominates: the SLOPE, for nearly two thirds. Improving the absorbance reading would gain almost nothing as long as the calibration curve stays what it is, which is the opposite of intuition, because the reading is what we have just taken and the curve dates from an hour ago.

c)

The dilution multiplied by 100.005.00=20.0\dfrac{100.00}{5.00}=20.0, so Criver=20.0×1.596=31.9C_{river}=20.0\times 1.596=31.9 mg/L. It adds its own relative uncertainties: 0.0155.00×100=0.30 %\dfrac{0.015}{5.00}\times 100=0.30\,\% and 0.08100.00×100=0.08 %\dfrac{0.08}{100.00}\times 100=0.08\,\%, giving 6.8+0.38=7.2 %6.8+0.38=7.2\,\% and ΔC=0.072×31.9=2.3\Delta C=0.072\times 31.9=2.3 mg/L. We publish C=(32±2)C=(32\pm 2) mg/L. The glassware added only 0.40.4 percentage points to 7.27.2: it is not the culprit, and that is a useful result, because it is the glassware that we fuss over by instinct.

d)

Because the Beer-Lambert law stops being linear well before A=1.20A=1.20. Above A≈1A\approx 1, stray light in the instrument, the association of coloured molecules with one another and the change in refractive index of the solution bend the relationship downward: the extended line UNDERESTIMATES the concentration, and all the more the further you go. So it is not only that we have left the calibrated interval, it is that the law itself has changed shape, and adding more concentrated standards would save nothing. The correct move: dilute the sample by a known factor, read again, and multiply. A one-in-five dilution would bring AA to about 0.240.24, right in the calibrated zone.

e)

0.0080.012×100=67 %\dfrac{0.008}{0.012}\times 100=67\,\%: the result is worthless. The absolute uncertainty on the absorbance does not depend on the signal, it is always 0.0080.008; so it is the RELATIVE uncertainty that explodes as the signal approaches zero. At the other end, above A=1A=1, it is linearity that is lost. In between, the relative precision is best near the middle of the range. The working rule that follows, and that is applied without thinking in the lab: bring the sample, by dilution or by concentration, to the MIDDLE of the calibration curve, typically around A=0.3A=0.3 to 0.70.7, instead of reading it where it happens to be.

0.511.522.530.10.20.30.40.50.6A read = 0.305C = 1.6 mg/LC (mg/L)A

Exercise 5: Linearization: it is the plot that decides the order

The disappearance of a coloured reagent AA is followed in the spectrophotometer, and the calibration lets us write the concentrations directly. The readings, in mol/L, are: t=0t=0 min, 0.05000.0500; t=20t=20, 0.03820.0382; t=40t=40, 0.03150.0315; t=60t=60, 0.02620.0262; t=80t=80, 0.02290.0229; t=100t=100, 0.01990.0199.

The figure shows the raw plot of [A][A] against time. For the regressions, we are given Stt=7000S_{tt}=7000, then for y=ln⁡[A]y=\ln[A]: Sty=−63.26S_{ty}=-63.26 and Syy=0.5814S_{yy}=0.5814; and for y=1[A]y=\dfrac{1}{[A]}: Sty=2101.5S_{ty}=2101.5 and Syy=631.19S_{yy}=631.19.

204060801001200.010.020.030.040.050.06t (min)[A] (mol/L)
  • a) What can be concluded from the raw plot in the figure, and what cannot?
  • b) Calculate the coefficient of determination of the regression of ln⁡[A]\ln[A] against tt, then that of 1[A]\dfrac{1}{[A]} against tt. Which order does the plot point to?
  • c) Calculate the rate constant and give its unit. Check the initial concentration that the regression predicts.
  • d) First order nevertheless gives r2r^{2} above 0.980.98. The residuals of that regression are, in order: +0.061+0.061, −0.027-0.027, −0.039-0.039, −0.043-0.043, +0.004+0.004, +0.044+0.044. What do you read in them that the coefficient of determination does not say?
  • e) Calculate the first half-life from your results, then the time needed to fall to 0.01000.0100 mol/L.

Type your answers, the page tells you right or wrong 0/12

a)
b)
c)
d)
e)
Show the solution

Answers

  • a) It rules out zero order and nothing more: first and second order cannot be told apart on it
  • b) r2=0.983r^{2}=0.983 for first order against 0.99950.9995 for second order: the reaction is second order
  • c) k=0.300k=0.300 L/(mol⋅\cdotmin), and the regression gives back [A]0=0.0500[A]_{0}=0.0500 mol/L
  • d) An arc pattern, + − − − + ++\,-\,-\,-\,+\,+, that r2r^{2} cannot see
  • e) t1/2=66.6t_{1/2}=66.6 min and t=266t=266 min to fall to 0.01000.0100 mol/L
a)

We read a decay that flattens out, so a rate that falls as the reagent becomes scarce, and we can rule out zero order, whose integrated law [A]=[A]0−kt[A]=[A]_{0}-kt would give a descending STRAIGHT LINE. That is all. First order and second order both produce a curved, flattening decay, and by eye they cannot be told apart. This plot therefore settles nothing, and that is precisely why we linearize: we do not compare two curves, we compare two ALIGNMENTS, which a human eye can judge and a ruler can check.

b)

First order: r2=(−63.26)27000×0.5814=4001.84069.8=0.983r^{2}=\dfrac{(-63.26)^{2}}{7000\times 0.5814}=\dfrac{4001.8}{4069.8}=0.983. Second order: r2=(2101.5)27000×631.19=44163004418330=0.9995r^{2}=\dfrac{(2101.5)^{2}}{7000\times 631.19}=\dfrac{4416300}{4418330}=0.9995. The second is thirty times closer to 11 in terms of residual gap: 0.00050.0005 against 0.0170.017. The plot of 1[A]\dfrac{1}{[A]} against tt is the one that lines up, so the reaction is SECOND ORDER. The point of method: an r2r^{2} is never read alone, it is COMPARED with that of the competing model, and it is the gap from 11 that is compared, not the value itself.

c)

For second order, 1[A]=1[A]0+kt\dfrac{1}{[A]}=\dfrac{1}{[A]_{0}}+kt: the slope is +k+k. k=StyStt=2101.57000=0.300k=\dfrac{S_{ty}}{S_{tt}}=\dfrac{2101.5}{7000}=0.300 L/(mol⋅\cdotmin). The units: 1[A]\dfrac{1}{[A]} is in L/mol, divided by minutes, so L⋅\cdotmol−1⋅^{-1}\cdotmin−1^{-1}, which are indeed those of a second-order rate constant. The y-intercept is yˉ−ktˉ=35.00−0.300×50=19.99\bar y-k\bar t=35.00-0.300\times 50=19.99 L/mol, so [A]0=119.99=0.0500[A]_{0}=\dfrac{1}{19.99}=0.0500 mol/L: we recover the measured starting concentration, which confirms that the line was not forced.

d)

What you read is a PATTERN: the residuals are positive at both ends and negative in the middle, + − − − + ++\,-\,-\,-\,+\,+. Purely random scatter would change sign without any order; this one draws a curve. Taking residual as observed minus fitted, the first-order line therefore runs above the data in the middle and below it at the ends, which is the signature of a straight line fitted to curved points. The coefficient of determination sees none of this: it adds up squares and loses the sign, so it cannot tell six random deviations from six deviations arranged in an arc. Looking at the residuals takes thirty seconds and answers the question that r2r^{2} does not ask, the one about the MODEL.

e)

For second order, t1/2=1k[A]0=10.300×0.0500=66.6t_{1/2}=\dfrac{1}{k[A]_{0}}=\dfrac{1}{0.300\times 0.0500}=66.6 min, which is consistent with the table, where [A][A] is still 0.02620.0262 at 6060 min. To fall to 0.01000.0100 mol/L: t=1k(10.0100−10.0500)=100−200.300=266t=\dfrac{1}{k}\left(\dfrac{1}{0.0100}-\dfrac{1}{0.0500}\right)=\dfrac{100-20}{0.300}=266 min. Four times the first half-life does not divide the concentration by sixteen, far from it: a second-order reaction drags on endlessly, its rate falling as the square of what remains.

ln[A]t1/[A]tfirst order: r² = 0.983second order: r² = 0.9995

Part B: from the slope to the published constant (/50)

Exercise 6: The logarithm does not propagate like a concentration

An acetic acid solution is prepared at C=0.100C=0.100 mol/L, known to within 0.5 %0.5\,\%, and its pH is measured with an electrode calibrated on two buffers. The display reads pH=2.87\mathrm{pH}=2.87, and the manufacturer states ±0.02\pm 0.02 pH units.

Recall that for a weak acid HAHA of concentration CC, Ka=[H3O+]2C−[H3O+]K_{a}=\dfrac{[H_{3}O^{+}]^{2}}{C-[H_{3}O^{+}]}.

The diagram below shows the measurement.

2.87pH meterCH₃COOH(aq)0.100 mol/L ± 0.5 %± 0.02 unit
  • a) Calculate the hydronium ion concentration.
  • b) Calculate its relative uncertainty, then its absolute uncertainty. Compare with the relative uncertainty on the pH itself.
  • c) Calculate KaK_{a}, then its relative and absolute uncertainty. Write the publishable result.
  • d) Give pKa\mathrm{p}K_{a} with its uncertainty. The reference value is 1.8×10−51.8\times 10^{-5}: is the measurement compatible?
  • e) How many significant figures does the hydronium ion concentration deduced from pH=2.87\mathrm{pH}=2.87 carry?

Type your answers, the page tells you right or wrong 0/10

a)
b)
c)
d)
e)
Show the solution

Answers

  • a) [H3O+]=1.349×10−3[H_{3}O^{+}]=1.349\times 10^{-3} mol/L
  • b) 4.6 %4.6\,\%, that is 6.2×10−56.2\times 10^{-5} mol/L, against 0.70 %0.70\,\% read off the pH
  • c) Ka=(1.8±0.2)×10−5K_{a}=(1.8\pm 0.2)\times 10^{-5}, that is 9.7 %9.7\,\%
  • d) pKa=4.73±0.04\mathrm{p}K_{a}=4.73\pm 0.04, compatible with 4.744.74
  • e) Two, those of the mantissa
a)

[H3O+]=10−pH=10−2.87=1.349×10−3[H_{3}O^{+}]=10^{-\mathrm{pH}}=10^{-2.87}=1.349\times 10^{-3} mol/L.

b)

An uncertainty does not pass through a logarithm as a percentage. From pH=−log⁡[H3O+]\mathrm{pH}=-\log[H_{3}O^{+}] we get Δ[H3O+][H3O+]=ln⁡(10)×ΔpH=2.303×0.02=0.046\dfrac{\Delta[H_{3}O^{+}]}{[H_{3}O^{+}]}=\ln(10)\times\Delta\mathrm{pH}=2.303\times 0.02=0.046, that is 4.6 %4.6\,\%. Absolute uncertainty: 0.046×1.349×10−3=6.2×10−50.046\times 1.349\times 10^{-3}=6.2\times 10^{-5}, so [H3O+]=(1.35±0.06)×10−3[H_{3}O^{+}]=(1.35\pm 0.06)\times 10^{-3} mol/L. Yet the relative uncertainty on the pH is 0.022.87×100=0.70 %\dfrac{0.02}{2.87}\times 100=0.70\,\%: seven tenths of a percent on one side, nearly five percent on the other. This is the central trap of the chapter, and it fits in one sentence: a ΔpH\Delta\mathrm{pH} is an ABSOLUTE uncertainty on an exponent, hence a RELATIVE uncertainty on the concentration. The percentage read off the pH has no chemical meaning.

c)

Ka=(1.349×10−3)20.100−0.001349=1.820×10−60.09865=1.845×10−5K_{a}=\dfrac{(1.349\times 10^{-3})^{2}}{0.100-0.001349}=\dfrac{1.820\times 10^{-6}}{0.09865}=1.845\times 10^{-5}. The hydronium concentration appears SQUARED, so its relative uncertainty is doubled: 2×4.6=9.2 %2\times 4.6=9.2\,\%, plus the 0.5 %0.5\,\% of the prepared solution, giving 9.7 %9.7\,\%. Absolute uncertainty: 0.097×1.845×10−5=1.8×10−60.097\times 1.845\times 10^{-5}=1.8\times 10^{-6}. Publishable result: Ka=(1.8±0.2)×10−5K_{a}=(1.8\pm 0.2)\times 10^{-5}. Two figures on the pH gave two on the constant: the chain gained nothing and lost nothing, but the illusion of the calculator's four digits very nearly had us write 1.845×10−51.845\times 10^{-5}.

d)

pKa=−log⁡(1.845×10−5)=4.734\mathrm{p}K_{a}=-\log(1.845\times 10^{-5})=4.734. Going back into the logarithm divides the relative uncertainty by ln⁡10\ln 10: ΔpKa=0.0972.303=0.042\Delta\mathrm{p}K_{a}=\dfrac{0.097}{2.303}=0.042, so pKa=4.73±0.04\mathrm{p}K_{a}=4.73\pm 0.04. The reference 1.8×10−51.8\times 10^{-5} corresponds to pKa=4.74\mathrm{p}K_{a}=4.74, a deviation of 0.010.01, a quarter of the uncertainty: perfectly compatible. In relative terms on KaK_{a}, the deviation is 2.5 %2.5\,\% for an uncertainty of 9.7 %9.7\,\%, which says the same thing. Note in passing that ΔpKa=0.042\Delta\mathrm{p}K_{a}=0.042 is twice ΔpH=0.02\Delta\mathrm{pH}=0.02: the squared numerator comes through intact after the round trip through the logarithms.

e)

TWO. In a logarithm, only the MANTISSA, the part after the decimal point, carries significant figures; the characteristic, here the 22, merely gives the power of ten, exactly as the exponent of a scientific notation carries none. The mantissa 8787 has two digits, so [H3O+]=1.3×10−3[H_{3}O^{+}]=1.3\times 10^{-3} mol/L and not one digit more, which the 4.6 %4.6\,\% uncertainty confirms independently. The practical consequence is brutal: to obtain a concentration to three significant figures you would need a pH to three decimal places, 2.8702.870, that is, a pH meter and a calibration of a different class from that of a teaching laboratory.

Exercise 7: The exponent multiplies the uncertainty: a Ksp taken from a pH

Solid calcium hydroxide is stirred in distilled water until saturation, the mixture is filtered, and the pH of the filtrate is measured: pH=12.35±0.02\mathrm{pH}=12.35\pm 0.02 at 25 ∘25\,^{\circ}C. The dissolution is written Ca(OH)2(s)⇌Ca2++2 OH−Ca(OH)_{2}(s)\rightleftharpoons Ca^{2+}+2\,OH^{-}, so Ksp=4s3K_{sp}=4s^{3} where ss is the molar solubility.

The reference value at 25 ∘25\,^{\circ}C is Ksp=5.5×10−6K_{sp}=5.5\times 10^{-6}. Take Kw=1.0×10−14K_{w}=1.0\times 10^{-14}.

The diagram below shows the measurement.

12.35pH metersaturated filtrateof Ca(OH)₂, 25 °C± 0.02 unit
  • a) Calculate the hydroxide ion concentration, then the molar solubility.
  • b) Calculate the solubility product.
  • c) Give the relative uncertainty on the solubility, then that on the solubility product. Write the publishable result.
  • d) Calculate the relative deviation from the reference value, then express it as a number of uncertainties. Conclusion?
  • e) We want a solubility product to within 5 %5\,\%. What uncertainty on the pH would be needed? Would repeating the measurement ten times be enough?

Type your answers, the page tells you right or wrong 0/10

a)
b)
c)
d)
e)
Show the solution

Answers

  • a) [OH−]=2.239×10−2[OH^{-}]=2.239\times 10^{-2} mol/L and s=1.119×10−2s=1.119\times 10^{-2} mol/L
  • b) Ksp=5.61×10−6K_{sp}=5.61\times 10^{-6}
  • c) 4.6 %4.6\,\% on ss, 13.8 %13.8\,\% on KspK_{sp}, that is (5.6±0.8)×10−6(5.6\pm 0.8)\times 10^{-6}
  • d) 2.0 %2.0\,\% deviation, that is 0.140.14 uncertainty: fully compatible
  • e) ΔpH=0.007\Delta\mathrm{pH}=0.007, out of reach; repeating does not fix a calibration
a)

pOH=14.00−12.35=1.65\mathrm{pOH}=14.00-12.35=1.65, so [OH−]=10−1.65=2.239×10−2[OH^{-}]=10^{-1.65}=2.239\times 10^{-2} mol/L. Each formula unit dissolved releases TWO hydroxide ions, so s=[OH−]2=1.119×10−2s=\dfrac{[OH^{-}]}{2}=1.119\times 10^{-2} mol/L. The factor 22 here is the first trap, and it has nothing to do with uncertainties: it comes from the dissolution equation, which is written before any calculation.

b)

Ksp=4s3=4×(1.119×10−2)3=4×1.402×10−6=5.61×10−6K_{sp}=4s^{3}=4\times (1.119\times 10^{-2})^{3}=4\times 1.402\times 10^{-6}=5.61\times 10^{-6}.

c)

The solubility is half of [OH−][OH^{-}], and dividing by an exact number does not change the relative uncertainty: ln⁡(10)×0.02=4.6 %\ln(10)\times 0.02=4.6\,\%, the same as on [OH−][OH^{-}]. The solubility product, for its part, goes as s3s^{3}: a power MULTIPLIES the relative uncertainty by its exponent, so 3×4.6=13.8 %3\times 4.6=13.8\,\%. The 44 is exact and contributes nothing. Absolute uncertainty: 0.138×5.61×10−6=7.8×10−70.138\times 5.61\times 10^{-6}=7.8\times 10^{-7}. We publish Ksp=(5.6±0.8)×10−6K_{sp}=(5.6\pm 0.8)\times 10^{-6}. Two hundredths of a pH unit have become fourteen percent on the constant, through nothing more than the composition of a logarithm and a cube.

d)

∣5.61−5.50∣5.50×100=2.0 %\dfrac{|5.61-5.50|}{5.50}\times 100=2.0\,\%. In number of uncertainties: 0.11×10−60.78×10−6=0.14\dfrac{0.11\times 10^{-6}}{0.78\times 10^{-6}}=0.14. The deviation fits within a seventh of the uncertainty: the measurement is fully compatible with the reference, and nothing more can be asserted. It would be wrong to congratulate ourselves on the two percent as if it were a performance: with an error bar of fourteen percent, the experiment could just as well have come out at 5.05.0 or 6.2×10−66.2\times 10^{-6} with nothing to criticize.

e)

We would need 3×ln⁡(10)×ΔpH=0.053\times\ln(10)\times\Delta\mathrm{pH}=0.05, that is ΔpH=0.056.91=0.007\Delta\mathrm{pH}=\dfrac{0.05}{6.91}=0.007. No teaching-laboratory pH meter holds seven thousandths of a pH unit, if only because the calibration buffers themselves are given at ±0.01\pm 0.01. And repeating would change nothing: the ±0.02\pm 0.02 is the uncertainty of the electrode's CALIBRATION, hence a systematic error that shifts all the readings the same way. Repeating ten times would give ten almost identical values, all shifted alike, with the reassuring illusion of excellent precision thrown in. The only remedy is a change of method: titrating the filtrate with hydrochloric acid from a burette gives the same solubility with an uncertainty ten times smaller, because the burette passes through no logarithm.

Exercise 8: Five statements to correct

Each statement is to be declared true or false. If it is false, the expected correction is the correct sentence, not just the word changed.

  • a) “A coefficient of determination of 0.9980.998 proves that the chosen model is the right one.”
  • b) “The measured pH is 2.872.87: three digits written, so three significant figures, and the concentration is published with three figures.”
  • c) “A systematic error eventually disappears if the measurement is repeated enough times.”
  • d) “To find the order of a reaction, you always plot the concentration against time.”
  • e) “The relative uncertainty of a product is the sum of the ABSOLUTE uncertainties of its factors.”

Type your answers, the page tells you right or wrong 0/10

a)
b)
c)
d)
e)
Show the solution

Answers

  • a) False: it measures alignment, not the model, and it gives no uncertainty
  • b) False: two significant figures, so [H3O+]=1.3×10−3[H_{3}O^{+}]=1.3\times 10^{-3} mol/L
  • c) False: repeating reduces only random error
  • d) False: you plot the linearizations, the raw plot rules out only zero order
  • e) False: in a product it is the relative uncertainties that add
a)

FALSE. The coefficient of determination measures how well the points LINE UP around the fitted line, nothing else. Over a narrow interval, a slightly curved line easily gives 0.9980.998, and that is exactly what happens in exercise 5, where first order shows 0.9830.983 while being the wrong model. Two moves are worth more than reading a number: compare the r2r^{2} with that of the COMPETING model, comparing the gaps from 11 and not the values, and look at the RESIDUALS, where an arc pattern exposes the curvature that r2r^{2} cannot see. The correct sentence: a coefficient of determination close to 11 says the points are aligned, it says neither that the model is right nor by how much the slope can move.

b)

FALSE. In a logarithm, only the mantissa carries significant figures: the characteristic 22 places the power of ten, like the exponent of a scientific notation. The mantissa 8787 has two, so [H3O+]=1.3×10−3[H_{3}O^{+}]=1.3\times 10^{-3} mol/L. The uncertainty confirms it separately: ln⁡(10)×0.02=4.6 %\ln(10)\times 0.02=4.6\,\%, which already puts the second digit in question. The correct sentence: pH=2.87\mathrm{pH}=2.87 carries two significant figures, so the concentration is published with two figures.

c)

FALSE, and this is the costliest confusion in the laboratory. Repeating a measurement reduces the RANDOM part, which changes sign from one trial to the next and averages out. A systematic error always falls on the same side: an electrode calibrated on an expired buffer shifts every pH by 0.050.05 units, and a thousand measurements will give a thousand values shifted by 0.050.05, with a superb precision thrown in that inspires confidence. Only a calibration, a blank, a control sample or an independent second method reveals it. The correct sentence: repeating reduces random error and never touches systematic error.

d)

FALSE. The plot of [A][A] against tt settles nothing: it is a straight line only for zero order, and first and second order give two curved decays that the eye cannot tell apart. You plot instead the LINEARIZATIONS, ln⁡[A]\ln[A] against tt for first order and 1[A]\dfrac{1}{[A]} against tt for second order, and it is the one that lines up that names the order. The correct sentence: you plot the candidate linearizations and keep the one that gives a straight line, the raw plot only making it possible to rule out zero order.

e)

FALSE, twice over. The rule concerns the RELATIVE uncertainties, and for a product or a quotient it is they that add: ΔPP=Δxx+Δyy\dfrac{\Delta P}{P}=\dfrac{\Delta x}{x}+\dfrac{\Delta y}{y}. It is the ABSOLUTE ones that add in a sum or a difference. Adding absolute uncertainties in a product makes no dimensional sense anyway: you would be adding millilitres to grams. The correct sentence: for a product or a quotient, the relative uncertainties add; for a sum or a difference, the absolute ones do.

Exercise 9: Problem: assaying the iron in a tablet by spectrophotometry

A dietary supplement tablet states 27.027.0 mg of elemental iron. It is crushed, dissolved in dilute hydrochloric acid, filtered and made up in a 250.0250.0 mL volumetric flask (±0.15\pm 0.15 mL). A 5.005.00 mL aliquot is taken with the pipette (±0.015\pm 0.015 mL), the reducing agent and the o-phenanthroline are added, and the solution is made up in a second 250.0250.0 mL flask (±0.15\pm 0.15 mL). The cuvette gives A=0.412±0.008A=0.412\pm 0.008.

We use the calibration curve of the previous exercises: A=aCA=aC with a=(0.191±0.008)a=(0.191\pm 0.008) L/(mg⋅\cdotcm), established from 0.5000.500 to 2.5002.500 mg/L.

The diagram below shows the preparation chain.

flask 1: 250.0 mL± 0.15 mLtabletpipette5.00 mL± 0.015 mLflask 2: 250.0 mL± 0.15 mL+ reagentscuvetteA = 0.412± 0.008
  • a) Calculate the concentration in the cuvette, then that in the stock flask, then the mass of iron in the tablet.
  • b) Calculate the relative deviation from the value stated on the label.
  • c) Draw up the uncertainty budget, item by item, then give the publishable mass.
  • d) Which item dominates? Quantify the gain from a calibration curve twice as tight, then from perfect glassware. What should you conclude about how to improve this assay?
  • e) Name the health and safety precautions and the waste treatment this procedure requires.

Type your answers, the page tells you right or wrong 0/13

a)
b)
c)
d)
e)
Show the solution

Answers

  • a) 2.1572.157 mg/L in the cuvette, 107.9107.9 mg/L in the stock flask, m=26.97m=26.97 mg
  • b) 0.1 %0.1\,\%
  • c) 6.5 %6.5\,\% in total, so m=(27±2)m=(27\pm 2) mg: tablet complies
  • d) The slope, two thirds of the budget; 32 %32\,\% gain against 6 %6\,\% for the glassware
  • e) Glasses, lab coat, reducing agent in a fume hood, and waste to the heavy-metal container, never down the sink
a)

Cuvette: C=0.4120.191=2.157C=\dfrac{0.412}{0.191}=2.157 mg/L, a value right in the middle of the calibrated range, which is the sign of a procedure that was properly sized. Dilution factor: 250.05.00=50.0\dfrac{250.0}{5.00}=50.0, so the stock flask is at 50.0×2.157=107.950.0\times 2.157=107.9 mg/L. Mass in the 250.0250.0 mL: m=107.9×0.2500=26.97m=107.9\times 0.2500=26.97 mg. We keep this number unrounded for what follows and will round only once the uncertainty is known.

b)

∣26.97−27.0∣27.0×100=0.1 %\dfrac{|26.97-27.0|}{27.0}\times 100=0.1\,\%. A spectacular agreement, which we are about to see proves next to nothing.

c)

Five items, all as relative uncertainties since everything is a product or a quotient. Absorbance: 0.0080.412×100=1.9 %\dfrac{0.008}{0.412}\times 100=1.9\,\%. Slope of the calibration: 4.2 %4.2\,\%. 5.005.00 mL pipette: 0.0155.00×100=0.30 %\dfrac{0.015}{5.00}\times 100=0.30\,\%. Flask of the dilution: 0.15250.0×100=0.060 %\dfrac{0.15}{250.0}\times 100=0.060\,\%. Stock flask: 0.060 %0.060\,\%. Total 6.5 %6.5\,\%, so Δm=0.065×26.97=1.8\Delta m=0.065\times 26.97=1.8 mg, rounded to one significant figure: 22 mg. We publish m=(27±2)m=(27\pm 2) mg. The tablet complies, and that is all we are entitled to write: this assay could not tell a 2727 mg tablet from a 28.528.5 mg one. The 0.1 %0.1\,\% deviation of b) is therefore a happy coincidence, not a proof of accuracy.

d)

The slope dominates, with 4.24.2 points out of 6.56.5, nearly two thirds of the budget; the absorbance supplies 30 %30\,\% and all the glassware 6 %6\,\%. Halving the uncertainty on the slope, by doubling the number of standards and reading each of them three times, brings the total down to 4.5 %4.5\,\% and Δm\Delta m to 1.21.2 mg: a gain of 32 %32\,\%. Making the glassware PERFECT, that is, eliminating three items entirely, brings the total down to 6.1 %6.1\,\% and Δm\Delta m to 1.71.7 mg: a gain of 6 %6\,\%. The conclusion is counterintuitive and holds for this whole chapter: we instinctively fuss over the visible gesture, the mark on the flask and the meniscus in the pipette, while the result is governed by an invisible object that has already been put away, the calibration line. Improving the calibration pays five times more than improving the glassware.

e)

Safety glasses and a lab coat throughout the procedure, gloves for the acid work-up. Dilute hydrochloric acid is always poured into water and never the other way round, and the safety data sheet is read before the session, not after the accident: the reducing agent used to keep the iron in the +2 oxidation state, hydroxylamine hydrochloride or hydroquinone depending on the protocol, is toxic and is handled only with a spatula, in a fume hood for weighing. On the waste side, nothing goes down the sink: the final solution contains complexed iron, o-phenanthroline and the reducing agent, and it goes into the heavy-metal waste container, neutralized beforehand if it is still strongly acidic. The cuvette rinse water goes there too. The reason is not administrative: a group of thirty students produces several litres of iron solution per session, and o-phenanthroline is toxic to aquatic organisms.

Exercise 10: Problem: an activation energy read from a slope

The decomposition of hydrogen peroxide catalyzed by the iodide ion is followed in the spectrophotometer at five temperatures set by a thermostatted bath. The rate constants obtained are: 298298 K, k=9.2×10−3k=9.2\times 10^{-3}; 308308 K, 1.97×10−21.97\times 10^{-2}; 318318 K, 4.25×10−24.25\times 10^{-2}; 328328 K, 8.20×10−28.20\times 10^{-2}; 338338 K, 1.62×10−11.62\times 10^{-1}, all in s−1^{-1}.

We are given R=8.314R=8.314 J/(mol⋅\cdotK), and for the regression of ln⁡k\ln k against 1000T\dfrac{1000}{T}: Sxx=0.098587S_{xx}=0.098587, Sxy=−0.71157S_{xy}=-0.71157 and Syy=5.13696S_{yy}=5.13696. The literature reference value is Ea=58.8E_{a}=58.8 kJ/mol.

The figure shows the five measured constants as a function of temperature.

2902953003053103153203253303353403450.020.040.060.080.10.120.140.160.18T (K)k (1/s)
  • a) What is plotted on the x-axis and the y-axis, and what is the slope in terms of the activation energy?
  • b) Calculate the slope of the regression, then the activation energy in kilojoules per mole.
  • c) Calculate the coefficient of determination. A classmate concludes that the activation energy is known to the thousandth of a kilojoule. What do you reply?
  • d) The two extreme lines have slopes of −7.10-7.10 and −7.34-7.34. Give the activation energy with its uncertainty, then the deviation from the reference as a percentage and as a number of uncertainties.
  • e) Two defects are discovered after the fact: the thermostat read 1.01.0 K too low at every temperature, and the scratched cuvette inflated all the rate constants by 3 %3\,\%. Classify them, and say which one distorts the activation energy.

Type your answers, the page tells you right or wrong 0/14

a)
b)
c)
d)
e)
Show the solution

Answers

  • a) ln⁡k\ln k on the y-axis, 1T\dfrac{1}{T} in kelvins on the x-axis, slope −EaR-\dfrac{E_{a}}{R}
  • b) Slope −7.218-7.218, so Ea=60.0E_{a}=60.0 kJ/mol
  • c) r2=0.9998r^{2}=0.9998: the model holds, but no uncertainty comes out of it
  • d) Ea=(60.0±1.0)E_{a}=(60.0\pm 1.0) kJ/mol, deviation of 2.0 %2.0\,\%, that is 1.21.2 uncertainties
  • e) Both are systematic; the thermostat moves EaE_{a} by 0.6 %0.6\,\%, the scratched cuvette does not touch it
a)

We linearize k=Ae−Ea/RTk=Ae^{-E_{a}/RT} by taking the logarithm: ln⁡k=ln⁡A−EaR⋅1T\ln k=\ln A-\dfrac{E_{a}}{R}\cdot\dfrac{1}{T}. So we plot ln⁡k\ln k on the y-axis and 1T\dfrac{1}{T} on the x-axis, here in the form 1000T\dfrac{1000}{T} to handle numbers around 33 rather than thousandths. The slope is −Ea1000R-\dfrac{E_{a}}{1000R}, and the y-intercept is ln⁡A\ln A. The temperature must be in KELVINS, and this forgotten conversion is the most frequent fault of the calculation.

b)

a=SxySxx=−0.711570.098587=−7.218a=\dfrac{S_{xy}}{S_{xx}}=\dfrac{-0.71157}{0.098587}=-7.218. Hence Ea=−a×1000R=7.218×1000×8.314=6.00×104E_{a}=-a\times 1000R=7.218\times 1000\times 8.314=6.00\times 10^{4} J/mol, that is 60.060.0 kJ/mol, a typical value for a catalyzed reaction in solution. A useful check: the slope is necessarily NEGATIVE, since heating speeds things up; a positive slope would signal swapped axes or a forgotten sign on the exponent.

c)

r2=(−0.71157)20.098587×5.13696=0.506330.50641=0.9998r^{2}=\dfrac{(-0.71157)^{2}}{0.098587\times 5.13696}=\dfrac{0.50633}{0.50641}=0.9998. The reply to the classmate comes in two parts. First, the five points are indeed very well aligned, which confirms that the Arrhenius law describes this reaction well over this interval: that is information about the MODEL, and it is good to have. Second, it still says nothing about the uncertainty: a coefficient of determination cannot be converted into an error bar, and the five digits the calculator displays for the slope are the result of a division, not a measurement. A separate method is needed, and that is the subject of the next question.

d)

Ea,max=7.34×8.314=61.0E_{a,max}=7.34\times 8.314=61.0 kJ/mol and Ea,min=7.10×8.314=59.0E_{a,min}=7.10\times 8.314=59.0 kJ/mol, hence Ea=61.0+59.02=60.0E_{a}=\dfrac{61.0+59.0}{2}=60.0 kJ/mol and ΔEa=61.0−59.02=1.0\Delta E_{a}=\dfrac{61.0-59.0}{2}=1.0 kJ/mol. We publish Ea=(60.0±1.0)E_{a}=(60.0\pm 1.0) kJ/mol, that is 1.7 %1.7\,\%. Deviation from the reference: ∣60.0−58.8∣58.8×100=2.0 %\dfrac{|60.0-58.8|}{58.8}\times 100=2.0\,\%, that is 1.21.0=1.2\dfrac{1.2}{1.0}=1.2 uncertainties. The agreement is acceptable without being perfect: a deviation of a little over one uncertainty does not condemn the measurement, but it leaves open the possibility of a small bias, which the next question is about to examine.

e)

Both are SYSTEMATIC: they always fall on the same side, and no repetition would reveal them. They do not do the same damage, however. The thermostat shifts the five temperatures by the same amount, but the x-axis is in 1T\dfrac{1}{T}, where a constant shift is not transmitted as it stands: redoing the calculation with 299299, 309309, 319319, 329329 and 339339 K gives 60.460.4 kJ/mol, that is 0.6 %0.6\,\% more, about two and a half times less than the uncertainty. The scratched cuvette multiplies all the rate constants by 1.031.03, and so adds ln⁡(1.03)=0.0296\ln(1.03)=0.0296 to every ln⁡k\ln k: it translates the whole cloud upward without changing any gap, so the slope does not move by a digit and the activation energy is EXACTLY the same. Only the y-intercept is affected, hence the pre-exponential factor. Here is the thread of the whole set: a slope is blind to anything that shifts every point by the same amount, and that is precisely what makes a constant read from a slope more solid than a constant read off an instrument. The 2 %2\,\% deviation from the reference therefore comes from neither of these two causes.

2.933.13.23.33.4−5−4.5−4−3.5−3−2.5−2−1.5slope = -7.218Ea = 60.0 kJ/mol1000 / T (K⁻¹)ln k
Something missing from this set? A type of exercise your teacher gives, a problem you are stuck on, a mistake you spotted: tell me, I read every request.

Sending goes through your free space: that is how I can answer you and tell you when the exercise is added. An email is enough, no password. Create your space. Rather not create an account? Write to me directly.

What is it about?
I read every request myself, nobody else sees it. No need to write your name or your school.
When your request is dealt with, you hear about it by email and in your space.

By sending, you agree that I read your request to answer it. It is kept with your account, erased with it, and never published: privacy policy.

Next chapter Chemical kinetics, 202-SN2-RE

© Ahmed Squalli Houssaini. Exercise set published at www.letuteurscientifique.ca/en/exercices/cegep-chem-sn2-measurement-uncertainty. Free for personal and classroom use; republishing it elsewhere requires written permission (legal notice).