Electricity and Magnetism 203-SN2-RE • CEGEP, Montreal
Corrected exercises: lab measurements, uncertainties and graphical analysis (203-SN2-RE)
Here are fifteen corrected exercises on measurement and uncertainty in the electricity and magnetism lab, course 203-SN2-RE (formerly 203-NYB) of the revised Natural Science program (200.B1). The program prescribes this experimental component in all three physics courses, but it is practised on the setups of each one: here, the multimeter, the voltage divider, the real cell, the RC circuit, the solenoid and the Wheatstone bridge.
Part A lays down the rules on electrical quantities: a multimeter datasheet and its digits, the relative uncertainties of a quotient, series and parallel combinations, the extreme-values method and the differential method. Part B moves on to data analysis: repeated measurements, linear regression, extreme lines, semi-logarithmic linearization. Part C revisits what comes back from one lab to the next, and ends on the costliest trap of the chapter, the difference of two nearby quantities.
The thread of the set: in electricity, the uncertainty almost never comes from the hand that reads, it comes from the instrument and the circuit. A four-digit display does not have four correct digits, a voltmeter changes the circuit it measures, a diameter squared weighs double, and a small difference of two nearby voltages destroys the precision. So you work out BEFORE the lab what each quantity will contribute, and you improve the contribution that dominates, never “the experiment” in general.
Self-checking setType your answers and check them question by question: the page tells you right or wrong before you open the solution. With an account, every answer right on the first try earns points, and your right answers stay filled in when you come back.
•Digital multimeter: uncertainty ±(a%of reading+ndigits), one digit being one unit of the last displayed place. The digit term depends on the RANGE, not on the value measured.
•Writing a result: x=(xˉ±Δx) unit; the uncertainty to one significant figure, two if it starts with 1, and the value rounded to the same decimal place.
•Sum or difference: Δ(x±y)=Δx+Δy, ABSOLUTE uncertainties.
•Product, quotient, power: zΔz=∣a∣xΔx+∣b∣yΔy for z=kxayb, RELATIVE uncertainties. An exact factor, such as a counted number of turns, contributes nothing.
•Parallel: Rp=R1+R2R1R2 is not a product. You propagate with ∂R1∂Rp=(R1+R2R2)2; if the two relative tolerances are equal, the combination keeps them.
•Extreme-values method: Δz=2zmax−zmin, taking the HIGH bound of the quantities in the numerator and the LOW bound of those in the denominator.
•Repeated measurements: mean, standard deviation s=N−1∑(xi−xˉ)2, uncertainty of the mean Ns (type A). The datasheet uncertainty is type B, and repeating never reduces it.
•Random error: changes sign, averaging reduces it. Systematic error: same sign, averaging keeps it. The loading effect of a voltmeter and the resistance of the test leads are systematic.
•Linearizing an exponential: u=U0e−t/τ gives lnu=lnU0−τt, a line of slope −τ1.
•A component's tolerance and a measurement uncertainty are COMPARED, never added: two values are consistent when their intervals overlap.
Part A: the basics (/50)
Exercise 1: Reading a digital multimeter and writing a result
A digital multimeter has no reading uncertainty in the sense of a graduated ruler: the display is sharp. Its uncertainty is written in the DATASHEET, in the form ±(a%of reading+ndigits), where one digit is one unit of the last displayed figure.
The instrument used here is specified ±(0.5%+2digits) on DC voltage. On the 20.00 V range, it displays 12.47 V. The diagram below shows the display on the two ranges.
a) What is one digit worth on the 20.00 V range, in V?
b) Calculate the absolute uncertainty on the reading 12.47 V, then write the result, in V.
c) Calculate the relative uncertainty, in %.
d) On the 200.0 V range, the same instrument would display 12.5 V. Calculate the absolute and the relative uncertainty on that range.
e) What handling rule do these two results impose?
Show the solution
Answers
a)1 digit =0.01 V
b)ΔU=0.08 V, so U=(12.47±0.08) V
c)0.66%
d)ΔU=0.3 V, that is 2.1%
e)Always pick the smallest range that does not overload
a)
The 20.00 V range displays two decimals: the last figure is the hundredth of a volt, so ONE DIGIT IS WORTH 0.01 V. The word “digit” never means a significant figure, it means one unit of the last displayed place, and it changes with the range. It is the first thing to look for in a datasheet.
b)
The datasheet gives two terms to ADD. Proportional term: 0.5%×12.47=0.0624 V. Digit term: 2×0.01=0.02 V. Total: ΔU=0.0624+0.02=0.0824 V, which is rounded to ONE significant figure, 0.08 V. The value is aligned on the same decimal: U=(12.47±0.08) V. Writing (12.4700±0.0824) V would be wrong by excess of precision: the four digits of the display are not four correct digits.
c)
UΔU=12.470.0824=0.0066, that is 0.66%. A common multimeter sits between half a percent and two percent: this is the order of magnitude to keep in mind to decide whether a measurement is worth refining.
d)
On the 200.0 V range, one digit is worth 0.1 V. Proportional term: 0.5%×12.5=0.0625 V, almost unchanged. Digit term: 2×0.1=0.2 V, that is TEN TIMES more. Total 0.2625 V, rounded to 0.3 V, so U=(12.5±0.3) V and a relative uncertainty of 2.1%. The same voltage, the same instrument, and a measurement three times worse.
e)
The rule: ALWAYS STEP DOWN TO THE SMALLEST RANGE THAT DOES NOT SHOW OVERLOAD. The digit term is set by the range, not by the voltage measured; measuring 12 V on a 200 V range means paying the uncertainty of a 200 V measurement. This is the most common error of the first lab, and it leaves no visible trace in the notebook: the display looks just as sharp.
Tick the exercises you have done or want to review: a free account, no password, keeps your ticks and your right answers from one visit to the next, tells you which chapter to tackle next and lets you ask for the exercise you are missing. Create your space, an email is enough.
Exercise 2: Ohm's law: a quotient, so relative uncertainties
A resistance is determined by measuring at the same time the voltage across it and the current through it, then forming the quotient R=IU.
Measurements: U=(6.42±0.05) V and I=(28.4±0.3) mA.
In a PRODUCT or a QUOTIENT, it is the RELATIVE uncertainties that add, never the absolute ones.
a) Calculate the resistance, in Ω.
b) Calculate the relative uncertainty on the resistance, in %.
c) Calculate the absolute uncertainty and write the result, in Ω.
d) Which of the two measurements should be improved first? Justify with the numbers.
e) A student writes ΔR=ΔIΔU=0.00030.05≈167 Ω. Where is the flaw in the reasoning?
Show the solution
Answers
a)R≈226.1 Ω
b)1.84%
c)ΔR≈4 Ω, so R=(226±4) Ω
d)The current: it contributes 1.06% against 0.78% for the voltage
e)It divides two uncertainties; an uncertainty is not a quantity to plug into the formula
a)
First convert the current to amperes, I=28.4 mA =0.0284 A, otherwise the resistance comes out in kilo-ohms without you noticing. R=0.02846.42=226.06 Ω. We provisionally keep all the decimals: the rounding is decided at the end, once the uncertainty is known.
b)
Each factor contributes its relative uncertainty. For the voltage: 6.420.05=0.00779, that is 0.78%. For the current: 28.40.3=0.01056, that is 1.06%. The two add, whether the quantity is in the numerator or the denominator: RΔR=0.00779+0.01056=0.01835, that is 1.84%. The fact that I is in the denominator changes NOTHING about the sign: an uncertainty is never subtracted.
c)
ΔR=226.06×0.01835=4.15 Ω, rounded to 4 Ω. The value is aligned on the unit: R=(226±4) Ω. The decimals of 226.06 disappear, because an uncertainty of 4 Ω makes the tenths digit meaningless.
d)
The current, without hesitation: it contributes 1.06% against 0.78% for the voltage, that is 58% of the total. This is the reasoning you do BEFORE going back to the lab: stepping the ammeter down one range, or increasing the current, will almost halve the uncertainty, whereas a finer voltmeter would gain almost nothing.
e)
The student treats the uncertainties as if they were themselves a voltage and a current, and applies Ohm's law to them. An uncertainty is not a physical quantity of the circuit: it is the width of an interval. It propagates by a rule on RELATIVE uncertainties, never by substituting it into the formula. The result obtained, 167 Ω, is moreover of the same order as the resistance itself, which should have raised an alarm: an uncertainty of 74% on a measurement made with a multimeter is physically absurd.
Exercise 3: Series and parallel: two combinations, one relative uncertainty
Two carbon resistors carry the manufacturer's tolerance of 5%: R1=(100±5)Ω and R2=(220±11)Ω.
In SERIES, the equivalent resistance is a SUM, so the ABSOLUTE uncertainties add. In PARALLEL, Rp=R1+R2R1R2 is neither a sum nor a simple product: you propagate with partial derivatives, ∂R1∂Rp=(R1+R2R2)2.
a) Calculate the equivalent resistance in series and its absolute uncertainty, in Ω.
b) Calculate the relative uncertainty of the series combination, in %.
c) Calculate the equivalent resistance in parallel, in Ω.
d) Calculate the absolute, then the relative uncertainty of the parallel combination.
e) The result of d) is no accident. Prove that if R1ΔR1=R2ΔR2=t, then RpΔRp=t.
Show the solution
Answers
a)Rs=320 Ω and ΔRs=16 Ω
b)5.0%
c)Rp=68.75 Ω
d)ΔRp≈3.4 Ω, that is 5.0%
e)ΔRp=tR1+R2R1R2=tRp
a)
Rs=100+220=320 Ω. It is a SUM, so the absolute uncertainties add: ΔRs=5+11=16 Ω. Result: Rs=(320±16)Ω, which can be written (3.20±0.16)×102Ω if you want to keep only the useful figures.
b)
32016=0.050, that is 5.0%. Since both resistors have the same relative tolerance of 5%, the sum keeps it: this is a general property of sums of quantities of the SAME sign.
c)
Rp=100+220100×220=32022000=68.75 Ω. Order-of-magnitude check, to be made every time: an equivalent resistance in parallel is always SMALLER than the smaller of the two, so below 100 Ω, and here it is a little more than two thirds of 100 Ω. It is consistent.
d)
The two partial derivatives: (R1+R2R2)2=(320220)2=0.4727 and (R1+R2R1)2=(320100)2=0.0977. Hence ΔRp=0.4727×5+0.0977×11=2.363+1.074=3.44 Ω, and RpΔRp=68.753.44=0.050, again 5.0%. We write Rp=(69±3)Ω.
e)
With ΔR1=tR1 and ΔR2=tR2, we get ΔRp=(R1+R2)2R22tR1+(R1+R2)2R12tR2=(R1+R2)2tR1R2(R2+R1)=tR1+R2R1R2=tRp. In other words: when all the components of a circuit carry the SAME relative tolerance, any series or parallel combination keeps it. This is what makes the 5% tolerance usable directly on a whole network, without recalculating. The symmetric trap: a student who added the relative uncertainties for the parallel case would find 10%, double the truth.
Exercise 4: The extreme-values method on dissipated power
The power dissipated by the Joule effect in a resistor supplied under a known voltage is P=RU2.
Measurements: U=(12.0±0.2) V and R=(47±2)Ω.
The EXTREME-VALUES METHOD requires no rule to memorize: you recompute the quantity with the combinations of bounds that make it largest and then smallest, and take the half-range, ΔP=2Pmax−Pmin. The diagram below shows the circuit.
a) Calculate the power, in W.
b) To obtain Pmax, should you take R at its maximum or at its minimum? Justify.
c) Calculate Pmax and Pmin, in W.
d) Calculate ΔP and write the result, in W.
e) Find ΔP again with the relative-uncertainty rule, without forgetting the square, and compare.
Show the solution
Answers
a)P≈3.06 W
b)R at its MINIMUM, because R is in the denominator
c)Pmax≈3.308 W and Pmin≈2.842 W
d)ΔP≈0.23 W, so P=(3.1±0.2) W
e)PΔP=7.59% gives ΔP≈0.23 W: same result
a)
P=4712.02=47144=3.064 W.
b)
The power grows with U and DECREASES with R, which is in the denominator. To maximize P, you therefore take U at its maximum AND R at its MINIMUM: Umax=12.2 V with Rmin=45Ω. This is the one delicate point of the method, and the one people miss: taking both high bounds would give a value that is neither the maximum nor the minimum.
c)
Pmax=4512.22=45148.84=3.3076 W and Pmin=4911.82=49139.24=2.8416 W. We check in passing that the central value 3.064 W lies between the two, and roughly in the middle: if it did not, that would be a sign that a bound was badly chosen.
d)
ΔP=23.3076−2.8416=20.4660=0.2330 W, rounded to 0.2 W. Result: P=(3.1±0.2) W. The central value is rounded to the tenth, since the uncertainty is on the tenth.
e)
For P=RU2, the voltage appears SQUARED, so its relative uncertainty counts TWICE: PΔP=2UΔU+RΔR=2×12.00.2+472=0.0333+0.0426=0.0759, that is 7.59%. Hence ΔP=3.064×0.0759=0.233 W: the two methods give the same number to three decimals. They are not equivalent in general, the extreme-values method being exact and the relative-uncertainty method a first-order approximation, but as long as the relative uncertainties stay below a few percent, the gap is invisible. Forgetting the factor 2 of the square would drop the result to 5.9%, an underestimate of one quarter: this is the number-one trap of the chapter.
Exercise 5: The differential method on the field of a solenoid
A solenoid of N=500 turns and length L, carrying a current I, produces at its centre a field B=Lμ0NI, with μ0=4π×10−7 T·m/A.
Measurements: I=(2.40±0.05) A and L=(0.320±0.002) m. The number of turns is COUNTED, so it is exact.
The DIFFERENTIAL METHOD propagates each uncertainty through the corresponding partial derivative, then adds the contributions in absolute value: ΔB≈∂I∂BΔI+∂L∂BΔL. The diagram below shows the solenoid in cross-section.
a) Calculate the field at the centre, in mT.
b) Calculate the contribution of the current, ∂I∂BΔI, in mT.
c) Calculate the contribution of the length, ∂L∂BΔL, in mT.
d) Write the result with its uncertainty, in mT, then give the relative uncertainty in %.
e) Why does the number of turns contribute nothing, even though it appears in the formula?
Show the solution
Answers
a)B≈4.71 mT
b)0.098 mT
c)0.029 mT
d)B=(4.71±0.13) mT, that is 2.7%
e)N is a counted integer, with no uncertainty: ΔN=0
a)
B=0.3204π×10−7×500×2.40=4π×10−7×3750=4.712×10−3 T, that is 4.71 mT. The intermediate calculation LNI=0.320500×2.40=3750 A/m is the linear density of ampere-turns: it is this that sets the field, not the three quantities taken separately.
b)
∂I∂B=Lμ0N=0.3204π×10−7×500=1.9635×10−3 T/A. Contribution: 1.9635×10−3×0.05=9.82×10−5 T, that is 0.098 mT. Since B is proportional to I, you could also write directly IΔI×B=2.400.05×4.712=0.098 mT: the two readings are the same.
c)
∂L∂B=−L2μ0NI=−LB, whose absolute value is 0.3204.712×10−3=1.4726×10−2 T/m. Contribution: 1.4726×10−2×0.002=2.95×10−5 T, that is 0.029 mT. The MINUS SIGN disappears: an uncertainty is never subtracted, you add the contributions in absolute value, because you are looking for the worst case.
d)
ΔB=0.098+0.029=0.127 mT, rounded to 0.13 mT since the first significant figure is a 1 and you then keep two. Result: B=(4.71±0.13) mT, and BΔB=4.7120.127=2.7%. The current accounts for 2.1 percentage points and the length for 0.6: it is the ammeter that should be improved, not the measuring tape.
e)
Because ΔN=0. Counting 500 turns gives exactly 500, not 500±something: it is an INTEGER, not a measurement. The same reasoning holds for μ0, which is a defined constant, and for any exact factor, such as the 2 of a diameter converted to a radius. A quantity with no uncertainty does not contribute, even when it multiplies everything else. Conversely, if the turns were wound by hand and counted to within ±5, you would have to add 5005=1%.
Part B: problems and reasoning (/50)
Exercise 6: Repeated voltage measurements: when repeating stops helping
A laboratory power supply set to 5 V is read ten times in a row with the multimeter, on the 20.00 V range. The values, in volts, are:
Reminders: Uˉ=N∑Ui, experimental standard deviation s=N−1∑(Ui−Uˉ)2, uncertainty of the mean Ns. The instrument is still specified ±(0.5%+2digits).
The figure below places the ten readings.
a) Calculate the mean, in V.
b) Calculate the experimental standard deviation, in mV.
c) Calculate the uncertainty of the mean, in mV.
d) The display gives hundredths of a volt. Does the observed scatter come from the instrument or from the source? Give numbers.
e) The instrument uncertainty on this reading is 0.045 V. Should it be combined with the result of c)? What do you conclude about the value of repeating further?
Show the solution
Answers
a)Uˉ=5.023 V
b)s≈42 mV
c)10s≈13 mV
d)From the SOURCE: 42 mV against 5 mV of resolution, a factor of 8
e)Yes, and the instrument dominates: U=(5.02±0.06) V, repeating more gains nothing
a)
The sum is 50.23 V, so Uˉ=1050.23=5.023 V. We provisionally keep the thousandths, the mean of a sample legitimately having one more figure than the individual measurements.
b)
The deviations from the mean are +0.007, +0.047, −0.043, −0.013, +0.067, −0.063, +0.017, −0.023, +0.037, −0.033 V. Their sum is zero, which is the first check to make. The sum of squares is 1.601×10−2V2, so s=91.601×10−2=1.779×10−3=0.0422 V, that is 42 mV. The denominator is N−1=9, not 10: the mean has already used up one degree of freedom.
c)
Ns=100.0422=3.1620.0422=0.0133 V, that is 13 mV. This is the TYPE A uncertainty, the one that comes from the observed scatter, and it is three times smaller than the scatter of a single measurement.
d)
The display resolution is 0.01 V, so a single reading is already known to better than ±0.005 V from that alone. But the real scatter is 42 mV, EIGHT TIMES more. The ten readings are therefore not spread over one or two neighbouring digits: they sprawl over thirteen digits. It is the SOURCE that fluctuates, not the instrument that hesitates. A student who wrote U=(5.03±0.005) V after a single reading would have underestimated the uncertainty by a factor of eight.
e)
Yes, they must be combined: 13 mV is a type A uncertainty, 45 mV a TYPE B uncertainty, coming from the datasheet, and both are present at the same time. Adding them, ΔU=0.045+0.013=0.058 V, that is 0.06 V; combining them in quadrature, 0.0452+0.0132=0.047 V, that is 0.05 V. In both conventions, the instrument term LARGELY dominates. Practical conclusion: going from ten to forty measurements would divide 13 mV by two, so it would gain less than 7 mV out of a total of 58; it is not worth the time. Repeating a measurement only reduces the random part, and stops being useful as soon as the instrumental part exceeds it. To really gain, you need a better instrument, not more rows in the table.
Exercise 7: Linearizing Ohm's law: what five points say and one keeps quiet
The voltage across a resistor is recorded for five values of the current:
I (mA): 5.0, 10.0, 15.0, 20.0, 25.0. U (V): 1.12, 2.19, 3.31, 4.38, 5.52.
Linear regression of U against I gives a slope a and an intercept b.
a) Which quantity do you plot on the horizontal axis, which on the vertical axis, and what does the slope represent?
b) Calculate the slope of the regression, in Ω.
c) Calculate the intercept, in V. What should it equal, and what would a clearly nonzero value reveal?
d) The resistor is marked 220Ω. Calculate the relative difference between the slope and this value, in %.
e) A sixth point, I=40.0 mA, gives U=9.05 V. Does it line up? What should you conclude, and what would you have missed with a single measurement?
Show the solution
Answers
a)I on the horizontal axis, U on the vertical axis; the slope IS the resistance
b)a≈219.8 Ω
c)b≈0.007 V, so zero within rounding
d)0.09%
e)No: 8.80 V was predicted against 9.05 measured, the resistor is heating up
a)
You plot the CURRENT on the horizontal axis and the VOLTAGE on the vertical axis, because the current is what you impose and the voltage is what you read. Ohm's law is then written U=RI, of the form Y=aX, and the SLOPE IS THE RESISTANCE, in volts per ampere, that is in ohms. Be careful to convert milliamperes to amperes before calculating the slope: keeping the mA, you would get 0.22 and a resistance a thousand times too small.
b)
With Iˉ=0.015 A and Uˉ=3.304 V, we form ∑(Ii−Iˉ)(Ui−Uˉ)=5.495×10−2 and ∑(Ii−Iˉ)2=2.5×10−4, hence a=2.5×10−45.495×10−2=219.8Ω. Quick check without a calculator: the last point gives 0.0255.52=220.8Ω, of the right order.
c)
b=Uˉ−aIˉ=3.304−219.8×0.015=3.304−3.297=0.007 V. It SHOULD be zero, since with no current there is no voltage across a pure resistor. The 7 mV found is at the noise level of the measurements, hence consistent with zero. A clearly nonzero intercept, say 0.2 V, would signal something else in the circuit: a zero offset of the voltmeter, a stray electromotive force, or a diode in series.
d)
220∣219.8−220∣=2200.2=0.0009, that is 0.09%. This is well within the 5% tolerance of the component: the measurement confirms the marking, and confirms it a hundred times better than the tolerance required.
e)
The line predicts U=219.8×0.040+0.007=8.80 V. We measure 9.05 V, that is 0.25 V more, a gap of 2.9%, well beyond the noise of the first five points. The point DOES NOT LINE UP. The cause is physical: at 40 mA, the resistor dissipates P=UI=0.36 W, it heats up, and the resistance of a metallic conductor RISES with temperature; the local ratio is 0.0409.05=226Ω. The lesson is the one of the chapter: a single measurement would have given 226Ω without the slightest warning signal, and it would have been written down as the value of the resistance. It is the GRAPH, and only the graph, that shows the drop-off and lets you say from which current Ohm's law stops being linear. An isolated point can never contradict a model: it has nothing to be compared with.
Exercise 8: Five statements to correct
Each of the following five sentences was overheard in an electricity lab. Say what is wrong with it, then write the correct version. A correction that merely says “that is false” earns no points: you must name the rule that is broken.
a) “A multimeter that displays four digits gives four reliable significant figures.”
b) “For two resistors in parallel, you add the relative uncertainties, as in a product.”
c) “The ammeter is connected in series, so it does not change the current it measures.”
d) “Linear regression gives the slope to the thousandth, so the uncertainty on the resistance is negligible.”
e) “The measurement R=(226±4)Ω is inconsistent with the marked value 220Ω, since 226=220.”
Show the solution
Answers
a)False: the display says nothing about accuracy, only the datasheet does
b)False: the parallel formula is not a product; here the relative uncertainty stays 5%, not 10%
c)False: its internal resistance lowers the current, it is a systematic error
d)False: the precision of the calculation is not the uncertainty of the measurement
e)False: the measured interval [222,230] misses 220, but the 5% tolerance on the marking reconciles them
a)
FALSE. The number of digits displayed is a feature of the DISPLAY ELECTRONICS, not of the accuracy. An instrument specified ±(0.5%+2) displaying 12.47 V is known to ±0.08 V: the last displayed digit is already uncertain. Correct version: “the number of significant figures of a result is decided from the uncertainty, calculated from the datasheet, and it is almost always smaller than the number of digits displayed”.
b)
FALSE, and doubly so. First, the formula Rp=R1+R2R1R2 is not a product: it contains a sum in the denominator, so the relative-uncertainty rule does not apply as it stands. Second, the result is numerically wrong: for (100±5)Ω and (220±11)Ω, adding the relative uncertainties would give 10% whereas the exact calculation gives 5%, the DOUBLE of the truth. Correct version: “you propagate with partial derivatives; when both resistors have the same relative tolerance, the parallel combination keeps it”.
c)
FALSE. A real ammeter has an internal resistance, small but not zero, and inserting it in series INCREASES the total resistance, so it LOWERS the current. The error is larger the lower the resistance of the circuit. It is not a random error: it always has the same sign and repeating does not wipe it out. Correct version: “an ammeter disturbs the circuit; you choose one whose resistance is much smaller than that of the circuit, and you correct for it when that is not possible”.
d)
FALSE. The number of decimals a calculator outputs has no connection with the uncertainty: it is a precision of CALCULATION, not an accuracy of MEASUREMENT. The slope inherits the uncertainty of the points, and you estimate it by the half-range of the slopes of the EXTREME LINES, or by the formula for the uncertainty on the slope. Correct version: “the regression gives the best slope, not its uncertainty; that is read from the scatter of the points around the line”.
e)
FALSE. Two different numbers can perfectly well designate the same quantity: that is precisely what the uncertainty allows. Here, the measurement covers [222,230]Ω and it indeed does not include 220. But the marked value is not an EXACT value: it carries the manufacturer's tolerance of 5%, hence [209,231]Ω. The two intervals overlap widely, so the values are CONSISTENT. Correct version: “two values are consistent if their intervals overlap; comparing the central numbers decides nothing”. This is the costliest reasoning error of the chapter, because it leads to declaring an experiment failed when it succeeded.
For c): a real ammeter is an ideal ammeter in series with its internal resistance rA. Inserting it adds rA to the resistance of the circuit, so it lowers the current it claims to measure, always in the same direction.
Exercise 9: Problem: the electromotive force and internal resistance of a cell
A cell supplies a variable resistor. For five settings, the current delivered and the voltage across the cell are recorded:
I (A): 0.10, 0.20, 0.30, 0.40, 0.50. U (V): 1.52, 1.47, 1.43, 1.38, 1.34.
The model of the real cell is U=ε−rI: the terminal voltage is the electromotive force (emf) minus the internal drop.
a) Why does the terminal voltage decrease when the current increases? What do you then read on the slope and on the intercept?
b) Calculate the slope of the regression, then the internal resistance, in Ω.
c) Calculate the electromotive force, in V.
d) The two extreme lines through all the uncertainty bars have slopes of −0.42 and −0.48 V/A. Give r with its uncertainty.
e) The label announces 1.5 V. With Δε=0.02 V, is the measurement consistent? What should you conclude about the “nominal value”?
Show the solution
Answers
a)The internal drop rI grows with the current; slope =−r, intercept =ε
b)Slope −0.45 V/A, so r=0.45 Ω
c)ε≈1.563 V
d)r=(0.45±0.03) Ω
e)Not consistent, and that is normal: a nominal value is not an expected value
a)
The current also flows through the INSIDE of the cell, which has a resistance r; it loses the voltage rI there. What remains available at the terminals is therefore ε−rI, and it decreases linearly with the current. The relation has the form Y=b+aX with X=I and Y=U: the SLOPE is −r, so the internal resistance is the OPPOSITE of the slope, and the INTERCEPT is ε, the voltage you would read at zero current, that is with the cell disconnected.
b)
Iˉ=0.30 A and Uˉ=1.428 V. We form ∑(Ii−Iˉ)(Ui−Uˉ)=−0.045 and ∑(Ii−Iˉ)2=0.10, hence a=0.10−0.045=−0.45 V/A. The internal resistance is r=−a=0.45Ω. The negative sign of the slope is NOT an error: it carries the physics, and taking it positive would give a cell that gets stronger when you draw on it.
c)
ε=Uˉ−aIˉ=1.428−(−0.45)×0.30=1.428+0.135=1.563 V. Notice that you do NOT get it by measuring the open-circuit voltage with a voltmeter: the voltmeter itself draws a small current. Extrapolating to zero current is the right method, and it is one of the rare cases where the intercept carries the sought result rather than a check.
d)
Δr=2∣−0.48∣−∣−0.42∣=20.48−0.42=0.03Ω. So r=(0.45±0.03)Ω, a relative uncertainty of 6.7%. Extreme lines are the standard graphical tool: you draw the steepest and the shallowest lines that still pass through all the uncertainty bars, and the half-range of their slopes gives the uncertainty.
e)
The interval of the measurement is [1.54,1.58] V, and 1.5 V does not belong to it: strictly speaking, the measurement and the label are INCONSISTENT. That does not mean the experiment failed. A nominal cell value is not an expected value in the metrological sense: it is a category name, like the 220Ω of a resistor, and a new alkaline cell commonly delivers 1.55 to 1.60 V, the value falling toward 1.5 V and then below as it discharges. The correct conclusion is therefore: “the measured electromotive force, (1.563±0.020) V, is that of a new cell, above its nominal value, which is expected”. Comparing a measurement with a commercial label without knowing what the label guarantees is the most common error of judgment in lab reports.
Exercise 10: Problem: the time constant of an RC circuit by semi-logarithmic regression
A capacitor charged under 9.00 V discharges through a resistor R=47 kΩ known to 5%. The voltage across it is recorded:
t (s): 0, 5, 10, 15, 20. u (V): 9.00, 6.02, 4.04, 2.70, 1.81.
The discharge follows u(t)=U0e−t/τ, which is not a straight line. You LINEARIZE it by taking the logarithm.
a) Show that lnu is an affine function of t, and say what its slope and its intercept equal.
b) Calculate the slope of the regression of lnu against t, in s−1.
c) Deduce the time constant, in s.
d) Calculate the capacitance, in µF. The capacitor is marked 270 µF at ±20%: is this consistent?
e) You could have timed the half-discharge instead. Calculate it, then say what the five-point method adds.
Show the solution
Answers
a)lnu=lnU0−τt: slope −τ1, intercept lnU0
b)a≈−0.0802 s−1
c)τ≈12.5 s
d)C≈265 µF, consistent with [216,324] µF
e)t1/2≈8.6 s; the five points prove in addition that the decay IS exponential
a)
Taking the natural logarithm of both sides of u=U0e−t/τ: lnu=lnU0+ln(e−t/τ)=lnU0−τt. It indeed has the form Y=b+aX with Y=lnu and X=t: the SLOPE is −τ1 and the INTERCEPT is lnU0. So you do not plot u against t, which would give a curve from which nothing precise can be read, but lnu against t.
b)
The logarithms are 2.1972, 1.7951, 1.3962, 0.9932, 0.5933. The successive differences, −0.402, −0.399, −0.403, −0.400, are almost identical: this is the signature of a true exponential, and it is a check to make before any calculation. The regression gives tˉ=10 s, lnu=1.3950, ∑(ti−tˉ)(lnui−lnu)=−20.048, ∑(ti−tˉ)2=250, hence a=250−20.048=−0.0802 s−1.
c)
τ=−a1=0.08021=12.47 s, about 12.5 s. Check with the intercept: b=1.3950+0.0802×10=2.197, and e2.197=9.00 V, exactly the announced initial voltage. When this check fails, it means that the initial voltage was misread or that the stopwatch did not start with the discharge.
d)
C=Rτ=4700012.47=2.653×10−4 F, that is 265 µF. With 5% on the resistance and about 2.4% on the time constant, the relative uncertainty on C reaches 7.4%, so C=(265±20) µF. The marking 270 µF at ±20% covers [216,324] µF, which contains the measurement by a wide margin: CONSISTENT. Electrolytic capacitors have very wide tolerances, often −20% to +50%: finding 265 instead of 270 is no feat, and a gap of 15% would not have been alarming either.
e)
The half-discharge time is t1/2=τln2=12.47×0.693=8.64 s. Timing it is quicker, and for a simple estimate it is enough. What it does not give: the PROOF that the decay is really exponential. A single timing would give a number whatever the phenomenon, even if the capacitor leaks, if the resistor heats up or if the voltmeter drains current; the alignment of the five points on the semi-logarithmic line, on the other hand, checks it. It is the same argument as in exercise 7: an isolated point can never contradict a model. We add that an upward curvature of the semi-logarithmic plot signals a parallel leak, and a downward curvature signals a resistance that increases.
Part C: the classics (/50)
Exercise 11: Random or systematic: only one of the two is cured by repeating
A RANDOM error changes sign from one measurement to the next; the mean reduces it, and the standard deviation measures it. A SYSTEMATIC error always keeps the same sign; the mean does not touch it, and no standard deviation reveals it.
A voltage divider is set up: two equal resistors in series on a 12.0 V supply, and a voltmeter of internal resistance RV=1.00 MΩ across the second one.
a) Without the voltmeter, what voltage is across the second resistor, in V?
b) With R=100 kΩ, calculate the resistance seen by the divider, then the voltage actually read, in V.
c) Calculate the absolute and the relative error made. Would repeating the measurement a hundred times make them disappear?
d) The circuit is rebuilt with R=1.0 kΩ. Calculate the voltage read and the relative error, in %.
e) Classify as random or systematic: the 60 Hz mains noise, the resistance of the test leads, the rounding of the last digit, the thermal drift of a standard resistor.
Show the solution
Answers
a)6.00 V
b)90.9 kΩ in parallel, and 5.71 V read
c)−0.29 V, that is −4.8%; no, it is systematic
d)6.00 V read, error −0.05%
e)Random: the mains noise, the rounding. Systematic: the leads, the thermal drift
a)
Two equal resistors in series share the voltage equally: U=212.0=6.00 V. This is the value we want to measure, and the one an ideal voltmeter, of infinite resistance, would display.
b)
The voltmeter is connected in PARALLEL with the second resistor, so the divider no longer sees 100 kΩ but 100+1000100×1000=1100100000=90.9 kΩ. The voltage read becomes 12.0×100+90.990.9=12.0×0.4762=5.71 V. The voltmeter has not measured the voltage of the circuit: it has measured the voltage of the circuit AS MODIFIED BY IT.
c)
Absolute error: 5.71−6.00=−0.29 V. Relative error: 6.00−0.29=−4.8%. Repeating a hundred times would change NOTHING: every measurement would give the same low value, the mean would still be 5.71 V, and the standard deviation would be tiny, giving the illusion of an excellent measurement. This is exactly the danger of a systematic error: it disguises itself as precision. No statistical treatment detects it; only analysis of the CIRCUIT reveals it.
d)
With R=1.0 kΩ, the parallel combination is 1000+1061000×106=999.0Ω, and the voltage read is 12.0×1999.0999.0=5.997 V, an error of −0.05%, a hundred times smaller. The rule: the loading effect is negligible as long as RV≫R, typically by a factor of a thousand. Symmetrically, an ammeter must have a resistance much SMALLER than that of the circuit. When that is not possible, as here with 100 kΩ resistors, you do not change instruments: you CORRECT the result by the calculation of b).
e)
The 60 Hz mains noise is RANDOM: it makes the reading fluctuate on both sides, and the mean reduces it. The rounding to the last digit is RANDOM too, as long as the true value does not systematically fall at the same place between two digits. The resistance of the test leads is SYSTEMATIC: it always adds in the same direction, and it is corrected by shorting the probes to read the offset, then subtracting it. The thermal drift of a standard resistor is SYSTEMATIC over the course of a session: the resistance rises steadily while the setup warms up, which biases all the measurements in the same direction; you fight it by letting the setup stabilize, or by alternating the order of the measurements.
For b): the voltmeter behaves like its internal resistance RV connected in parallel with the second resistor. The bottom of the divider is no longer 100 kΩ but 90.9 kΩ (RV=1.00 MΩ), and the voltage read falls to 5.71 V instead of 6.00 V.
Exercise 12: The colour code, the tolerance and the zero of the ohmmeter
A resistor carries the bands red, red, brown, gold. The code: the first two bands give the digits, the third the multiplier, the fourth the tolerance, gold being ±5%.
The TOLERANCE is not a measurement uncertainty: it is a manufacturer's guarantee on the scatter of its production. The two are compared, they are not added.
The diagram below shows the resistor and its bands.
a) Give the nominal value, the tolerance in ohms, and the interval guaranteed by the manufacturer.
b) The ohmmeter gives (226±4)Ω. Is the component within specification?
c) A second resistor from the same batch gives (238±3)Ω. Conclusion?
d) The ohmmeter probes are shorted: it displays (0.6±0.1)Ω. What should be done with this reading, and does the verdict of c) change?
e) A student adds the tolerance and the measurement uncertainty to obtain ±15Ω. Why is this a flaw in the reasoning?
Show the solution
Answers
a)220Ω, ±11Ω, so [209,231]Ω
b)Yes: [222,230] lies entirely within [209,231]
c)No: [235,241] does not meet [209,231]
d)Subtract it: 237.4Ω, the verdict does not change
e)It mixes a manufacturing guarantee with a measurement uncertainty
a)
Red is 2, red is 2, brown as the multiplier is ×10: the nominal value is 22×10=220Ω. The gold band gives ±5%, that is 0.05×220=11Ω. The manufacturer therefore guarantees [209,231]Ω, and nothing more: it does not promise 220.0Ω.
b)
The measurement covers [222,230]Ω. This interval is entirely contained in the guaranteed interval: the component is WITHIN SPECIFICATION. Note that the nominal value 220 does not need to belong to the measurement interval, and indeed it does not here. The question asked is “has the manufacturer kept its promise”, not “does the measurement land exactly on the label”.
c)
The measurement covers [235,241]Ω, which does NOT meet [209,231]: the two intervals are disjoint, with 4Ω between the manufacturer's upper bound and the lower bound of the measurement. Two explanations remain open: the component is out of tolerance, or the ohmmeter is biased. As long as the second has not been ruled out, we have no right to blame the first.
d)
These 0.6Ω are the resistance of the LEADS and contacts: they add to every reading, always in the same direction, so it is a systematic error that is corrected by subtraction. The corrected measurement is 238−0.6=237.4Ω, with an uncertainty raised to 3+0.1=3.1Ω, so [234.3,240.5]Ω. The interval still does not meet [209,231]: the verdict DOES NOT CHANGE, the resistor is indeed out of tolerance. The correction was still worth making: on a 2Ω resistor, the same 0.6Ω would have meant 30% error.
e)
Because the two numbers do not describe the same thing. The tolerance describes the SCATTER OF A POPULATION of components out of the factory; the uncertainty describes what OUR MEASUREMENT does not know about THIS component. Adding them amounts to confusing “I do not know which of the resistors in the drawer I picked up” with “I do not know exactly what the one I am holding is worth”. Concretely, the rule is: you CONFRONT the two intervals, you compare them, you never combine them. The only thing you add to a measurement uncertainty is another measurement uncertainty, like the 0.1Ω of question d).
For d): the leads and contacts form a small resistance in series with what you measure. With the probes joined, the ohmmeter reads only that, 0.6 Ω; on the resistor, it reads R+0.6 Ω, which you correct by subtracting: 237.4 Ω.
Exercise 13: The slide-wire Wheatstone bridge: why a null method wins
A slide-wire bridge compares an unknown resistance with a standard. A uniform wire of 100.0 cm is stretched along a ruler; a slider divides it into two sections of lengths ℓ and 100.0−ℓ, whose resistances are proportional to the lengths.
At balance, the central galvanometer shows no current and Rx=Rstd100.0−ℓℓ.
Data: Rstd=100.0Ω known to 0.1%, balance obtained for ℓ=(68.4±0.3) cm.
a) Calculate the unknown resistance, in Ω.
b) Calculate the relative uncertainty contributed by the two lengths, in %.
c) Add the contribution of the standard, then write the result with its uncertainty, in Ω.
d) A direct ohmmeter gives (219±7)Ω. Are the two measurements consistent, and which one do you keep?
e) Where should the slider be placed for the relative uncertainty to be as small as possible? What practical consequence for the choice of the standard?
Show the solution
Answers
a)Rx≈216.5Ω
b)1.39%
c)1.49% in total, so Rx=(216±3)Ω
d)Consistent; keep the bridge, twice as precise
e)At the MIDDLE of the wire, ℓ=50.0 cm: choose a standard close to Rx
a)
Rx=100.0×100.0−68.468.4=100.0×31.668.4=216.46Ω. Since the slider is beyond the middle, the ratio exceeds 1 and the unknown resistance is larger than the standard: an immediate qualitative check.
b)
The formula is a quotient of two lengths, so the relative uncertainties add. Mind the delicate point: the two lengths are NOT independent, but both carry the uncertainty Δℓ=0.3 cm, that of the slider position. 68.40.3=0.00439, that is 0.44%, and 31.60.3=0.00949, that is 0.95%. Total: 1.39%. The SHORT section dominates, because the same absolute uncertainty weighs relatively more there.
c)
The standard contributes 0.1%, so RxΔRx=1.39+0.10=1.49% and ΔRx=216.46×0.0149=3.2Ω, rounded to 3Ω. Result: Rx=(216±3)Ω. The standard hardly contributes: that is the whole point of paying for a good one, it drops out of the uncertainty budget.
d)
The intervals are [213,219]Ω for the bridge and [212,226]Ω for the ohmmeter. They overlap widely, so the two measurements are CONSISTENT. You keep the BRIDGE: its relative uncertainty is 1.5% against 3.2% for the ohmmeter, twice as good. The deeper reason is that the galvanometer only serves to detect a ZERO: neither its calibration, nor its linearity, nor the voltage of the cell enters the result, which depends only on a ratio of lengths and on a standard. A null method turns a measurement problem into a comparison problem, and a comparison is always done better than an absolute reading.
e)
The relative uncertainty is 0.3(ℓ1+100−ℓ1), a sum that is MINIMAL when the two terms are equal, that is for ℓ=50.0 cm: it then falls to 0.3×(501+501)=0.012, that is 1.2% instead of 1.39%. And it blows up toward the ends: at ℓ=90 cm, it equals 0.3×(901+101)=3.3%, almost three times worse. Practical consequence, and this is the real lesson of the exercise: YOU CHOOSE THE STANDARD AS CLOSE AS POSSIBLE TO THE UNKNOWN RESISTANCE, so that the balance falls near the middle of the wire. A 10Ω standard to measure 216Ω would give a balance near 96 cm and a mediocre measurement, even though the standard itself would be excellent.
Exercise 14: The resistivity of a wire: the diameter counts twice
The resistance of a uniform wire is R=SρL, where S=4πd2 is the area of the cross-section. From it you get the resistivity, ρ=LRS=4LπRd2.
Measurements on a nichrome wire: R=(12.81±0.05)Ω, L=(1.500±0.002) m, and the diameter with a micrometer screw gauge, d=(0.405±0.005) mm. The diagram below shows the wire, not to scale.
a) Calculate the area of the cross-section, in m².
b) Calculate the resistivity, in Ω·m.
c) Give the three relative contributions. Which one dominates, and why?
d) Calculate the absolute uncertainty and write the result. Nichrome is tabulated at 1.10×10−6 Ω·m: conclusion?
e) The micrometer is replaced by a vernier caliper, Δd=0.05 mm. What does the relative uncertainty become, in %?
Show the solution
Answers
a)S≈1.288×10−7 m²
b)ρ≈1.100×10−6 Ω·m
c)0.39%, 2.47% and 0.13%: the diameter dominates
d)ρ=(1.10±0.03)×10−6 Ω·m, consistent with the table
e)25%: the measurement is no longer worth anything
a)
First convert the diameter to metres, d=0.405 mm =4.05×10−4 m. S=4πd2=4π×(4.05×10−4)2=4π×1.6403×10−7=1.288×10−7 m². Forgetting the conversion is the most common fault here, and it goes unnoticed: the final result then comes out a million times too large, which looks like nothing in the tables.
The resistance appears to the power 1, the length to the power −1, the diameter to the power 2. So: RΔR=12.810.05=0.39%; 2dΔd=2×0.4050.005=2×1.23=2.47%; LΔL=1.5000.002=0.13%. The DIAMETER dominates, for two reasons that compound: it is measured on a tiny quantity, so its relative uncertainty is already the largest of the three, and it enters SQUARED, which doubles it again. By itself it makes up 83% of the uncertainty budget.
d)
Total: 0.39+2.47+0.13=2.99%, so Δρ=1.100×10−6×0.0299=3.3×10−8 Ω·m, rounded to 0.03×10−6. Result: ρ=(1.10±0.03)×10−6 Ω·m. The interval [1.07,1.13]×10−6 contains the tabulated value 1.10×10−6 Ω·m: the measurement is CONSISTENT with the table, and the wire is indeed nichrome.
e)
With Δd=0.05 mm, the contribution of the diameter rises to 2×0.4050.05=24.7%, and the total to 25.2%. The resistivity would become (1.1±0.3)×10−6 Ω·m: an interval wide enough to include nichrome as well as constantan or manganin, hence a measurement that no longer lets you identify the metal. Changing the instrument on the ONE dominant quantity takes the measurement from 3% to 25%, without any other measurement having moved. This is the clearest illustration of the rule of the chapter: you do not improve “the experiment”, you improve the contribution that dominates, and you spot it by calculation BEFORE going into the lab.
Exercise 15: Problem: the same internal resistance by difference, and the factor of four lost
We return to the cell of exercise 9, but with the quick method: measure the open-circuit voltage, then the loaded voltage for a known current, and form r=IUopen−Uloaded.
Measurements: Uopen=(1.565±0.005) V, Uloaded=(1.520±0.005) V, I=(0.100±0.005) A.
a) Calculate the internal resistance, in Ω.
b) Calculate the absolute, then the relative uncertainty on the DIFFERENCE of the two voltages.
c) Calculate the uncertainty on r and write the result, in Ω.
d) Compare with the (0.45±0.03)Ω obtained by regression. By what factor is the quick method worse?
e) The measurement is repeated at I=(0.500±0.005) A, where the loaded voltage falls to (1.340±0.005) V. What happens to the uncertainty? State the general rule.
Show the solution
Answers
a)r=0.450Ω
b)Δ(ΔU)=0.010 V, that is 22.2%
c)27.2% in total, so r=(0.45±0.12)Ω
d)Four times worse
e)r=(0.45±0.03)Ω: you make the difference grow, never the reverse
a)
ΔU=1.565−1.520=0.045 V, so r=0.1000.045=0.450Ω. The central value is exactly the one given by the regression of exercise 9: the two methods do not contradict each other, they differ in PRECISION, which is quite another matter.
b)
It is a DIFFERENCE, so the absolute uncertainties ADD: Δ(ΔU)=0.005+0.005=0.010 V. And this is where everything is decided: this 10 mV uncertainty is on a gap of only 45 mV, a relative uncertainty of 0.0450.010=22.2%. Each of the two voltages, on its own, was known only to 0.3%. The subtraction destroyed the information: it kept the uncertainty and removed the quantity.
c)
The current contributes 0.1000.005=5.0%. Total: 22.2+5.0=27.2%, so Δr=0.450×0.272=0.122Ω, rounded to 0.12Ω. Result: r=(0.45±0.12)Ω, meaning that the internal resistance is somewhere between 0.33 and 0.57Ω.
d)
The regression gave ±0.03Ω, that is 6.7%; the quick method gives ±0.12Ω, that is 27.2%. It is FOUR TIMES worse, for two reasons that compound: it rests on a small difference of two nearby quantities, and it uses only two points where the regression uses five. The time saved is real, but it is paid for dearly.
e)
At 0.500 A, the difference becomes 1.565−1.340=0.225 V, five times larger, while its absolute uncertainty stays 0.010 V: the relative uncertainty drops to 0.2250.010=4.4%. The current now contributes 0.5000.005=1.0%, so 5.4% in total and Δr=0.450×0.054=0.024, that is r=(0.45±0.03)Ω: the quick method catches up with the regression. THE GENERAL RULE: a difference of two nearby quantities is the worst place to put a measurement, because the absolute uncertainty survives intact while the quantity itself vanishes. Three remedies, in order: make the difference GROW by changing the conditions, as here by drawing five times more current; measure the difference DIRECTLY when the instrument allows it, for example with a differential setup; or replace the two points by a regression on several points, which amounts to measuring a slope rather than a gap.
Something missing from this set?A type of exercise your teacher gives, a problem you are stuck on, a mistake you spotted: tell me, I read every request.
Sending goes through your free space: that is how I can answer you and tell you when the exercise is added. An email is enough, no password. Create your space.
Rather not create an account? Write to me directly.